Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for
power
supply lines
not applica-
ble
not applicable
(for INTERNALLY
POWERED ME
EQUIPMENT)
Surge
IEC 61000-4-5
± 1 kV line(s)
to line(s)
not applica-
ble
not applicable
(for INTERNALLY
POWERED ME
EQUIPMENT)
Voltage dips,
short interrup-
tions and volt-
age variations
on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in
U
T
)
for 0.5 cycle
40% U
T
(60% dip in
U
T
)
for 5 cycles
70% U
T
(30% dip in
U
T
)
for 25 cycles
<5% U
T
(>95% dip in
U
T
)
for 5 sec
not applica-
ble
not applicable
(For INTERNAL-
LY POWERED ME
EQUIPMENT
Power frequency
(50Hz/60Hz)
magnetic field
IEC 61000-4-8
10V/m
10V/m
Power frequency mag-
netic fields should be
at levels characteristic
of a typical location in
a typical commercial
or hospital environ-
ment.