Ceevac 

 High Vacuum High Flow 

Portable Suction Pump 

SUC 81030 

12 VDC | 100-240 V~  50/60 Hz

 

User Manual 

CLEMENTS

 

Manual No. SUC 81030 008

 

Issue 2

 

Summary of Contents for SUC 81030

Page 1: ...Ceevac High Vacuum High Flow Portable Suction Pump SUC 81030 12 VDC 100 240 V 50 60 Hz User Manual CLEMENTS Manual No SUC 81030 008 Issue 2...

Page 2: ...by authorised personnel in accordance with the instructions as described in this manual Operated in this way the Ceevac High Suction Pump will provide the standard of service specified Due to continu...

Page 3: ...and Operation 10 Using FLOVAC Disposable Liners 11 Battery Status and Charging 13 Cleaning 15 Spare Parts 17 Troubleshooting 18 Wiring Diagram 19 Diagnostic Setup 20 Return for Repair 20 Periodic Safe...

Page 4: ...ly trained and authorized personnel Contraindications Before using the Ceevac consult the instructions for use Failure to follow instructions in this manual could cause harm Do not use Ceevac for thor...

Page 5: ...rning time Minimum 5 minutes Power Pack Battery Charge External Powers pump Monitors and charges battery when mains power supply is connected Pump Piston type Filter Disposable hydrophobic and bacteri...

Page 6: ...g table In addition for vehicular transportation the unit should be kept upright State Parameter Minimum Maximum Operating Temperature 5 C 35 C Humidity 10 RH 93 RH Barometric Pressure 700 hPa 1060 hP...

Page 7: ...tion cannula Insulation Class II double insulation Hz Mains power frequency Direct current Battery On Off CE Mark in conformity with CE MEDDEV directive 93 42 CEE and subsequent changes Manufacturer R...

Page 8: ...rk Degree of protection against intrusion by body parts or objects an against ingress by water 1st Digit Penetration of Solids 2 Protected against solids 12 5 mm 2nd Digit Penetration of Liquids 1 Pro...

Page 9: ...ack charger 2 Confirm that electrical rating of power pack matches that of the mains power Avoid the use of power boards and extensions 3 If power pack charger is used maintain clear access to the mai...

Page 10: ...let 3 Connect the long silicone tubing to the jar PATIENT port 4 Select power source 1 Connect plug of supplied power pack charger to power socket at rear of pump OR 2 Connect plug of supplied cigaret...

Page 11: ...iners 1 Remove liner from package and extend liner from lid 2 Insert liner bag 1 into the matching size support jar 2 Press edge of lid firmly all around edge to ensure seal 3 Close the TANDEM inlet 3...

Page 12: ...e may be signs of decay and replacement is recommended Silicone Tubing The number of cleaning and sterilization cycles of the silicone tubing is dependent on the actual usage Tubing should be checked...

Page 13: ...battery charge Battery Charging Use only the supplied power pack charger model UE60 140429SPA1 Recharge the battery after each use to maintain the pump s maximum standby capacity When the battery is...

Page 14: ...ts Battery Service Life The following chart shows the expected service life of the battery against the depth of discharge In summary maximum service life of the battery will be obtained from a usage p...

Page 15: ...rilised with sterilants determined by local protocols and compatible with silicone tubing Suction tubing may be autoclaved to a maximum of 121 C for 15 minutes At higher temperatures the suction tubin...

Page 16: ...scoloured Also check the tubing nipples Check the seal ring on the lid and replace if hard cracked or perished Check the fit of the lid in the collection jar canister Check the overflow cut off valve...

Page 17: ...er Pack 1 SUC 81030 014 PCBA Electronic Control for Ceevac SUC 81030 015 Ba8ery SLA 12V 4 0Ah for Ceevac SUC 81030 016 Adapter Vehicle 12V Fused for Ceevac SUC 81030 017 Switch ON OFF for Ceevac SUC 8...

Page 18: ...the Control Panel indicates zero If not release vacuum and switch pump off and on 2 Pump Running But No Vacuum Check that all fittings are connected tightly Turn the vacuum controller knob on the cont...

Page 19: ...CLEMENTS CLEMENTS CLEMENTS CLEMENTS 19 Wiring Diagram Wiring Diagram...

Page 20: ...Disconnect all items from inlet and with pump switched off confirm that gauge reads zero A non zero reading indicates a faulty gauge Set vacuum control knob to maximum switch on pump and confirm that...

Page 21: ...indicates leaks in jar or connections 1 Occluded leak check Flow Check Max Vacuum Gauge Connect pump as shown in diagram Set vacuum control knob to maximum and switch on pump Connect jar inlet to cali...

Page 22: ...eturn for Repair Before returning the ACeevac for repair the external surfaces and any accessories must be carefully disinfected with a cloth soaked in methylated spirits or hypochlorite based solutio...

Page 23: ...voltage power supply network that supplies buildings used for domestic purposes EMC Information Tables per EN 60601 1 2 2014 In accordance with EN 60601 1 2 2014 Medical electrical equipment Part 1 2...

Page 24: ...put output lines The device doesn t change its state Mains power quality should be that of a typical commercial environment or hospital Surge EN 61000 4 5 1 kV differential mode The device doesn t cha...

Page 25: ...the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol Radiated Immunity EN 61000 4 3 10 V m 80 MHz to 2 7 GHz for non life s...

Page 26: ...ansmitter W 150 kHz to 80 MHz d 3 5 V1 P 80 MHz to 800 MHz d 12 E1 P 800 MHz to 2 7 GHz d 23 E1 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters with...

Page 27: ...CLEMENTS CLEMENTS CLEMENTS CLEMENTS 27 Notes Notes...

Page 28: ...CLEMENTS CLEMENTS CLEMENTS CLEMENTS 28 Notes Notes...

Page 29: ...CLEMENTS CLEMENTS CLEMENTS CLEMENTS 29 Notes Notes...

Page 30: ...ble to the original purchaser of the product 3 The product must not have had its serial number removed defaced or changed its casing opened its power supply altered or have been tampered with in any o...

Page 31: ...rms of the sale of the product or otherwise in connection with the product is limited to the amount paid by the purchaser to ICU Medical Australia for the product 10 Apart from any liability imposed b...

Page 32: ......