background image

 

CMS-2 CARBON MONOXIDE MONITOR 

Page 11 

 

 

 

 

 

© 2018 CLEMCO INDUSTRIES CORP.  

  www.clemcoindustries.com  

  Manual No. 22925, Rev. I, 04/18 

4.9.2

  Follow the steps in Section 4.6 Calibration 

Testing, except use impurity-free test gas in place of 10 
ppm gas. The alarm light will either remain

 red 

or turn 

green

. Proceed as follows: 

 

4.9.3

 

Alarm light turns 

green

 

 

4.9.3.1

 If the monitor is correctly calibrated and in 

working order, the alarm light will turn 

green

 during this 

test. This means the monitor was reading contaminated 
air. DO NOT USE THE COMPRESSED AIR FOR 
BREATHING. 

 

 

 

WARNING 

Do not breathe air that this instrument identifies 
as toxic until the source of contamination is 
identified and corrected. Breathing toxic gases 
can cause death. 

 

 

4.9.3.2

 Identify the source of contamination. Do not 

overlook the possibility that contaminated air entered the 
compressor intake. The contamination could be from 
engine or other exhaust entering the intake. If the source 
of contamination is temporary, the monitor will return to 
a non-alarm (safe) condition after the contamination is 
cleared from the compressed-air system. 
 

4.9.4

 

Alarm light remains red

 

 

4.9.4.1

  If the monitor is out of calibration or not in good 

working order, the alarm light will remain red.

 

 
4.9.4.2

  Calibrate the monitor per Section 4.7 and apply 

impurity-free gas per Section 4.9. If the alarm light 
remains red, the monitor requires service. Refer to 
Section 6.0.

 

 

 

 

 

WARNING 

Do not use compressed air monitored by this 
instrument for breathing unless the instrument 
is in good working condition. Using a monitor 
that is not in calibration or not working 
correctly can permit undetected CO to enter the 
breathing-air lines. Breathing toxic gases can 
cause death. 

 

 
 

 
 

5.0 GENERAL 

MAINTENANCE 

 

5.1 

Care and Handling 

 

5.1.1

 

This monitor is designed for portable, field use 

and is not adversely affected by normal handling that is 
required of any test and measurement instrument. 
 

5.1.2

  Transport the instrument in the passenger 

compartment of the service vehicle. 

 

NOTICE 

Do not subject this instrument to extreme heat or 
cold. Placing the instrument on the dashboard of 
the service vehicle in direct sunlight or similar 
conditions will elevate the temperature, which 
will damage the instrument’s electronic 
components. Temperatures at 0

F and below or 

case temperature at 125

F and above will 

damage the instruments components. 

 

5.2 Cleaning 

 

5.2.1

  The need to open the case periodically to 

calibrate the instrument exposes it to potential external 
contamination. Take care not to introduce contaminants 
into the instrument when the case is open.  
 

5.2.2

 

Clean the exterior of the case with a solution of 

water and mild detergent. Do not use solvent cleaners. 
 

5.3 

Calibration Test Schedule 

 

5.3.1

 

Test the calibration when it is initially set up and 

again the day after. Check it once a week for the first 
month. Check it at least once a month thereafter. Refer 
to Section 4.0 for test procedure. 
 

5.4 Calibration 

Schedule

 

 

5.4.1 

Avoid the urge to calibrate the instrument. 

Calibrate only when the calibration test shows it is 
required. See Section 4.0. 
 

5.5 Alarm 

Tests 

 

5.5.1

 

Although uncommon, alarm lights and horns do 

fail. Check their function before each use by placing the 
Run/Calibration toggle toward CALIBRATION. The 
external alarm light will immediately turn 

yellow

, and 

within a few seconds the intermittent audible alarm will 
sound. Placing the toggle in RUN position returns the 
alarms to their normal operation. 

Never use the 

respirator without first verifying that the monitor is 
in the operating mode; the exterior alarm light must 
be 

green

 with no audible alarm. 

 

 

 

Summary of Contents for CMS-2

Page 1: ...e WARNINGS are included for the health and safety of the operator and those in the immediate vicinity Electronic files include a Preface containing the same important information as in the orange cover WARNING 2018 CLEMCO INDUSTRIES CORP One Cable Car Dr Washington MO 63090 Phone 636 239 4300 Fax 800 726 7559 Email info clemcoindustries com www clemcoindustries com ...

Page 2: ...tion that if not avoided could result in minor or moderate injury WARNING Warning indicates a hazardous situation that if not avoided could result in death or serious injury DANGER Danger indicates a hazardous situation that if not avoided will result in death or serious injury 1 3 Table of Contents SUBJECT SECTION LOCATION INTRODUCTION 1 0 Scope of Manual 1 1 Safety Alerts 1 2 Table of Contents 1...

Page 3: ...ier 1 5 Operating Principles 1 5 1 The CMS monitor samples respiratory air from a breathing air source This enables one monitor to detect CO and other toxic gases in the entire breathing air system not just for one respirator The air to the respirator does not actually pass through the monitor WARNING When the CMS is correctly installed and maintained it monitors the level of CO and other oxidizab...

Page 4: ...ion Do not use the compressed air for breathing unless the exterior light is illuminated green Yellow Light Intermittent Alarm Horn approximately every 10 seconds Sample Air Failure This alerts the user that pressure in the test chamber is either too low or too high Low pressure is usually due to loss of pressure in the sample air air line High pressure is an indication that the internal regulator...

Page 5: ...d to seal pipe fittings The monitor will respond to the gases given off by these compounds Use Teflon tape for pipe thread sealer Do not use any silicone based lubricants up stream of this instrument Silicone exposure will prematurely damage the sensor 2 3 1 To ensure accurate and rapid air monitoring the monitor must be placed as close to the compressed air source as practical Ten feet of 1 8 ID ...

Page 6: ... AC to DC field conversion Serial No 4446 and greater To convert monitors with serial numbers lower than 4446 request Manual No 22925 Revision E or earlier The AC power cord provided with the monitor may be used using black as the positive A user supplied DC 2 wire cord may be used using red as positive If a new 12 volt cord is used splice the positive through the fuse and use the spade on the out...

Page 7: ...o operate relays for external alarms and shut down devices 2 6 Changing Mode Select Dip Switch to Delay Alarm Figure 5 NOTE In areas where intermittent radio frequency transmissions cause false alarms the monitor can be set to have a 10 second alarm delay to avoid false alarm conditions The dip switch is mounted on the circuit board as shown in Figure 5 WARNING Do not change the switch positions u...

Page 8: ...Section 3 3 3 2 4 If the instrument goes into an alarm exterior light remains red with continuous audible alarm after the warm up reinitiate warm up by terminating and restarting power 3 2 5 Continued alarm condition after several warm up periods indicates that the sample air exceeds the permissible contamination level or the instrument requires calibration Refer to Section 4 0 WARNING DO NOT use ...

Page 9: ...river when calibration is required 4 3 Stabilization Period 4 3 1 The monitor must be in operation for at least 30 minutes before calibration tests DO NOT test the instrument until it has gone through the warm up period and has been in operation with sample air flowing through it see Section 3 0 for at least 30 minutes 4 4 Prepare Calibration Connector NOTE Unlike ambient air calibration test gas ...

Page 10: ...un Calibration toggle toward CALIBRATION The external alarm light will immediately turn yellow and within a few seconds the intermittent audible alarm will sound the calibration indicator remains green 4 6 3 With the flow meter vertical refer to Figure 9 and slowly open the flow control knob to introduce test gas Test gas is entering the unit when the flow meter ball rises If the ball does not ris...

Page 11: ...r light turns red 4 7 5 Once the indicator is red turn the potentiometer in the opposite direction until the light turns green Then slowly turn the potentiometer from green to red several times to find the trip point The monitor is calibrated at the spot where the light turns red Return the instrument to operating mode per Section 4 8 4 8 Return Monitor to Operating Mode 4 8 1 Turn the flow contro...

Page 12: ...death 5 0 GENERAL MAINTENANCE 5 1 Care and Handling 5 1 1 This monitor is designed for portable field use and is not adversely affected by normal handling that is required of any test and measurement instrument 5 1 2 Transport the instrument in the passenger compartment of the service vehicle NOTICE Do not subject this instrument to extreme heat or cold Placing the instrument on the dashboard of t...

Page 13: ...b as the new sensor must be installed in the same orientation While supporting the circuit board carefully pull straight up to remove the sensor 6 1 9 If the sensor comes off with the housing note the tab location and remove the sensor by pulling on a couple of the sensor pins 6 1 10 Discard the old sensor to avoid attempts to reuse 6 1 11 Handle the new sensor taking care not to distort the pins ...

Page 14: ...sengage as soon as the pressure is corrected After the pressure is correctly set close the faceplate 7 2 4 If the flow meter ball does not raise or stays very low in the meter 7 2 4 1 Check internal instrumentation lines for breaks kinks or disconnection If a line has come loose reconnecting it will re establish function and the monitor can be correctly used NOTE This type of failure could be the ...

Page 15: ... 120 volt AC 22892 12 volt DC 25072 120 volt AC monitors can be field converted to 12 volt DC Refer to Section 2 4 1 Calibration connector assembly includes tube humidifier and flow control valve with connector 22893 2 Test gas 10 ppm 22865 Impurity free 11132 3 Tubing clear for Item 1 and internal per ft specify length required 29261 4 Regulator filter M5 22920 5 Filter element replacement for It...

Reviews: