background image

13/11/2018 

 

Page 1 

powerPRO Series 
Power Supplies 

Instruction Manual 

Catalogue Numbers

powerPRO300 

powerPRO500 

powerPRO3AMP 

Record the following for your records: 

Model _____________________ 

Catalogue No. _____________________ 

Date of Delivery _____________________ 

Warranty Period _____________________ 

Serial No. _____________________ 

Invoice No. _____________________ 

Purchase Order No. _____________________ 

Summary of Contents for powerPRO Series

Page 1: ...POWERPRO...

Page 2: ......

Page 3: ...PRO500 powerPRO3AMP Record the following for your records Model _____________________ Catalogue No _____________________ Date of Delivery _____________________ Warranty Period _____________________ Se...

Page 4: ...uipment Operation 5 Symbols 5 Potential Risk and Preventive Measures 6 Risk assessment table 6 Preventative measures 6 Packing List 7 Specifications 8 Operating Instructions 10 Installation 10 Control...

Page 5: ...ore operating this equipment 1 Read and follow the manual instructions carefully 2 Do not alter the equipment Failure to follow these directions could result in personal and or laboratory hazards as w...

Page 6: ...ation of the unit The powerPRO Power Supply has been designed to utilize shielded wires thus minimizing any potential shock hazard to the user Cleaver Scientific recommends against the use of unshield...

Page 7: ...uction to see if the proposed method will damage the equipment Equipment Operation Follow the guidelines below to ensure safe operation of the unit 1 NEVER access dangerous chemicals or other material...

Page 8: ...500 V Max current 800mA Max watt 300W Potential Risk and Preventive Measures Risk assessment table Frequency Risk Frequent Likely Possible Rare Unlikely Bruise Slash Electrical shock Power cord plug w...

Page 9: ...acking lists should be referred to as soon as the units are received to ensure that all components have been included The unit should be checked for damage when received Cleaver Scientific is liable f...

Page 10: ...be changed by user Program Storage 30 programmed files Program Multi Step Up to 6 steps Editable Program Function 1 Have typical running conditions program 2 Manual editable program Display 2 4 TFT Co...

Page 11: ...Furthermore the powerful specifications plus five terminal pairs can be used for multi electrophoresis units simultaneously Cleaver Scientific powerPRO Series Power Supply provides Constant Voltage or...

Page 12: ...the unit Press to move cursor left between parameters Pause Key Press to temporarily interrupt power to an operation in progress resume power after pausing without resetting the timer Press to move c...

Page 13: ...witch on the back 5 The screen will show a loading screen for few seconds then enter to the mode selecting page Constant Mode Use the Constant Voltage Current Power Operation for applications that req...

Page 14: ...lier will constantly operate until user manually stops it 5 Press the start Key to begin electrophoresis The LED next to the start key will light indicating the power supply is active Parameter values...

Page 15: ...he operation set the Program Mode as follow 1 Use the up and down arrow keys to select program mode then press the enter key to move to the next page 2 Use the up and down arrows to select the desired...

Page 16: ...run 8 Upon completion of the run an alarm will sound and the screen will display Complete 9 Press the start key to terminate the run and press the setting key to return to the settings screen Typical...

Page 17: ...gories that can be selected Choose the model according to your experiment need 3 After selecting model the pre set value will display These values can be adjusted as described previously 4 Press the s...

Page 18: ...chine back to local distributor or our Service Department for maintenance Operation stops with alarm The screen displays Electrophoresis leads are not connected to the power supply or to the electroph...

Page 19: ...arm Display shows Ground leak detected during run Check the electrophoresis system for improper grounding Restart the power supply by turning the Power switch off and on Operation stops with alarm Dis...

Page 20: ...nspect the fuses If the fuses are burned or there is a break in the fuses element replace the fuses with identical type of fuses 2 Place the fuses holder back into the compartment 3 Snap the cover clo...

Page 21: ...anyone other than CSL or an appointed distributor or representative are no longer under warranty from the time the unit was modified Units which have accessories or repaired parts not supplied by CSL...

Page 22: ......

Page 23: ......

Page 24: ......

Reviews: