P-02 Manual Temperature setting
After entering P-02, you can set the manual temperature by pressing . Long press
for speedy setting. Confirm your setting by pressing .
You can switch between °C and °F by pressing . Press to confirm. This setting will only
take effect after the meter restarted.
Manual Temperature setting range: -40.0-125.0
℃(
-40.0-257.0°F
)
Please refer to above description steps to set up P-02.
You can go to next parameter setting by pressing , or pressing key to quit and go
back to measurement mode.
Note:
1. Please make sure the Temperature Electrode is UNPLUGGED when manual Temperature
compensation. Otherwise, the unit will display “
Err
” and go back to the set up menu for fresh
operation.
2. It is not working to unplug the Temperature Electrode while setting. You have to set again after
removing the electrode and quit the current mode.
10