CA Pressure
Pressure
Pressure FM
Minimum air flow
170 lpm
170 lpm
Maximum air flow
400 lpm
250 lpm
Weight of unit
250 g
280 g
Inlet connection
Compatible with RECTUS series 25, 26 and CEYN320
Outlet connection
CA40x1/7“
Noise level of unit
61 dB
61 dB
Waist size
60 – 150 cm
60 – 150 cm
Supply pressure range
300 – 1000 kPa
400 – 1000 kPa
Recommended temperature range at work
+10 to +60 °C
+10 to +60 °C
Recommended air humidity range at workplace
20 % to 80 % Rh
20 % to 80 % Rh
Certification
EN 14594 Class 2A
EN 14594 Class 2A, 2B,
3A, 3B
10. Technical Data
Notified body for CE testing:
Výzkumný ústav bezpečnosti práce, v.v.i. – ZL
Testing Laboratory No. 1024, Jeruzalémská 9, 116 52 Praha 1
Authorized Body 235 | Notified Body 1024
6
The unit does not supply
the sufficient amount of air.
Air hose or air channel blocked.
Check and remove the possible ob-
stacle.
Air escapes through leakages.
Check all sealing elements and co-
nnections, check that the hose is not
damaged and without leaks.
The baffle is clogged.
Replace the baffle in CA Pressure .
The filter in CA Conditioner is clogged.
Replace the filter.
Fault
Probable cause
Recommendation
The unit does not work at
all.
Compressed air supply failure. Hose
damage.
Check the compressed air source.
Check that connecting hoses are not
damaged.
9. Possible Faults