OE 4002
MULTIMETER MEGOHMETRE AND INSULATION TESTER 1000V CAT. III
GENERAL SPECIFICATIONS
Complies with IEC/ EN 61010-1 1000V CAT III, 600 V CAT IV.
1000V DC Voltage, 750VAC Voltage (TRMS).
6600 Count (except for capacitance, temperature,insulation).
When the input terminals is V and COM, The overload protective voltage is up to AC 250 V (TRMS) ; When the
input terminals is mA and COM, protective Current is 0.4A; when the input terminals is Hi and Lo, The overload
protective voltage is 600V.
Batteries: Four AA batteries (NEDA 15A or IEC LR6) Meter use 1000 hours; Insulation test use: Meter can
perform at least 1000 insulation tests with fresh alkaline batteries at room temperature. These are standard tests
of 1000 V into 1 MΩ with a duty cycle of 5 seconds on and 25 seconds off.
Insulation Measurement Range: 0.01 MΩ to 2 GΩ.
Insulation Test Voltages: 50V, 100V, 250V, 500V, 1000V.
Insulation Source Voltage: + 20%, - 0%.
Insulation Short-Circuit Test Current : 1 .5mA nominal.
Insulation Auto Discharge : Discharge time <1 second for C = 1μF or less.
Insulation Live Circuit Detection : Inhibit test if terminal voltage > 20V prior to initialization of test.
Insulation Maximum Capacitive Load: Operable with up to 1μF load.
Storage Temperature : -40 °C to 60 °C.
Operating Temperature: 0°C to 40 °C .
Storage Altitude: 12000m.
Operating Altitude: 2000m 1000V CAT III, 3000m 1000V CAT II.
Temperature Coeffi cient: 0.05 × (specifi ed accuracy) per °C for temperatures < 18 °C or > 28 °C.
Relative Humidity: 40%~75%.
Size: 205(L)×102(W)× 58(H)mm.
Weight: approx 390g.