How to use Setup
25
Name
Each test sequence is given a name. This is to help the user when
selecting test sequences during automatic testing. When Name is
highlighted press the Enter button, then use the keyboard to enter a
new name for the Test Sequence (of up to 15 characters) and press
Enter to confirm.
Test Parameter Fields
By selecting one of the test parameter fields from the TEST
SETTINGS EDITOR, a list of options for that field will be displayed
as shown in the example below: -
When you first enter this menu all of the previous values and one of
the Test Parameter columns are highlighted. Use the cursor left and
right keys to highlight the Test Parameter to be changed and then
use the cursor up and down keys to highlight the required setting.
Use this process for each Test Parameter.
Use the Enter button or OK Fast key to save the settings and return
to the TEST SETTINGS EDITOR. Press the Red button to abort
changes made to the settings and return to the TEST SETTINGS
EDITOR.
----------------------LEAKAGE--------------------
O/P
LIMIT DUR
1
110V SKIP
2s
2
230V 0.25mA
5s
4
0.50mA 10s
5
0.75mA 30s 10
1.00mA 60s 15
1.50mA 120s
20
2.00mA 180s
25
UL 30
UL