Summary of Contents for N2 LCMS 1

Page 1: ...User Manual Nitrogen generator N2 LCMS 1...

Page 2: ...2 Sommario 1 Introduction 3 2 Safety 4 3 Description of the generator 6 4 Installation 8 5 Disassembly and transport 11 6 Use 12 7 Maintenance 17 8 Troubleshooting 22 9 Guarantee 23...

Page 3: ...he use of compressed air and ni trogen The CE mark guarantees that these generators have been designed and con structed in accordance with the following European directives Electromagnetic compatibili...

Page 4: ...hile it is connected to the electrical mains RISK OF FATAL INJURY BY ELECTROCUTION e Repairs and inspections must be carried out exclusively by SPECIALIST PER SONNEL in the event of faults impossible...

Page 5: ...stantly controlled by a safety valve that is activated without the need for electrical power source if pressure values exceed 11 8 bar 1180 kPa 2 4 Technical assistance The CLAIND technical assistance...

Page 6: ...chnical specifications 3 2 1 General 3 2 2 Electrical 3 2 3 Gas The degree of purity of the gas on output varies according to the type of use and application The data regarding the present model are s...

Page 7: ...d B PRESSURE GAUGE indicates the output pressure of the nitrogen C POWER KEY ON pushbutton D LOWER PANEL access to the filter zone for maintenance E CONNECTOR for electric power supply cable includes...

Page 8: ...of suitable equipment such as pallet trucks or lift trucks e The generator must always remain in a vertical position as it is not designed to be laid on its side 4 2 2 Packaging removal The generator...

Page 9: ...sed when laying the gas distribution lines CONNECTION Locate the nitrogen coupling G 1 4 female marked OUTLET at the rear of the generator Connect the line 4 4 Drain CONNECTION Locale the DRAIN outlet...

Page 10: ...or partially automatic e To ensure compliance with standards imposed by the Machinery directive the start up of the generator after electrical power supply is restored cannot normal ly be automatic a...

Page 11: ...ctric power cable Close all valves air inlet tanks and detach the pneumatic connections 5 2 Transport e Remember that during transport the generator must always remain in a vertical position If conser...

Page 12: ...n of the various keys DISPLAY The display comprises 2 lines each of 20 characters Usually the top line displays the functions or parameters and the lower line di splays the relative status and values...

Page 13: ...erator reaches the Standby status In this status the generator interrupts the normal production cycles and the display shows the message STAND BY 7 From this point the nitrogen is available for the li...

Page 14: ...ny time during production the generator can be stopped by pressing the key START STOP Depending on the phase activated the generator starts to depressurise imme diately or awaits conclusion of the pro...

Page 15: ...r status press the UP key The following page is displayed AIR N2 YYYY W WW X XX ZZZZ Where W WW air pressure on inlet to sieves expressed in bar X XX nitrogen pressure in the tank expressed in bar YYY...

Page 16: ...and holding the Function key for a few seconds The data will be displayed in sequence as shown in the table above To modify the value use arrow keys To acquire the set value press and hold key and pre...

Page 17: ...requests whether the maintenance has been performed with display of the following message SCHEDULED MAINTENANCE OF FILTER COM PLETE NO Press the up arrow key to select the answer YES or NO consecutive...

Page 18: ...n the ambient air Component Replacement frequency Compressor inlet filter 2 000 hours Coalescence filter 2 000 hours Silencer filters 2 000 hours General overhaul 4 000 hours REPLACEMENT Turn off the...

Page 19: ...ow Turn off the generator as described in par 6 5 Stop the generator and par 6 6 Generator shutdown Disconnect the power cable then re move the front lower panel by loose ning the fixing screws see fi...

Page 20: ...off the generator and disconnect the power cable remove the front lower panel by loosening the fixing screws Locate the position of the solenoid val ves at the lower left and right of the generator Lo...

Page 21: ...list The following table summarises the consumable parts available at CLAIND In the purchase order specify the model and serial number of the generator Description CONSUMABLE KIT FOR NG2301 LCMS1 200...

Page 22: ...itrogen tank falls below the set MINIMUM NITROGEN PRESSU RE value To shut off the alarm press Function key To exit the alarm conditions reduce to the nitrogen flow rate to the utility to ena ble press...

Page 23: ...costs for technicians to visit the client s premises The guarantee covers exclusively COSTS DERIVING FROM MANUFACTURING DEFECTS and does not include 1 Damage caused by negligence or improper use of th...

Page 24: ...Claind srl Via Regina 24 22016 Lenno CO Italy Tel 39 0344 56603 Fax 39 0344 56627 e mail service claind it UM LCMS1 en 2 fm...