Using High-Level Disinfectants with GUS

Step 1: Probe Soaking Instructions

CIVCO Medical Solution's new soak and rinse containers for the GUS G10VP and G14KA units are 17" tall. This
allows your probes to be disinfected past the handle and strain relief.

CAUTION: Older probes are not water-resistant above the handle and must not be soaked above the
top of the handle. Failure to follow these instructions may result in expensive probe damage, may lead
to unsafe use or inability to use, and may result in injury.

Check with your probe manufacturer to determine whether your probe’s handle or strain relief can be
submerged.

Every probe to be soaked MUST be marked on the cable as follows:

If your probe is NOT water-resistant:

Suspend the probe into an empty
container so that the top of the handle is
level with the top of the container.
Depress blue tab and place probe cable
into the grip clip. Mark the cable just
above the top of the grip clip using a
permanent marker or tape.

If your probe IS water-resistant:

Suspend

the

probe

into

an

empty

container so that the tip of the probe is
about 1/2" above the bottom of the
container.
Depress blue tab and place probe cable
into the grip clip using a permanent
marker or tape

IMPORTANT:

By keeping the red mark above the grip clip, the probe will always be at the right length.

DISCLAIMER:

CIVCO MEDICAL SOLUTIONS IS NOT RESPONSIBLE FOR ANY DAMAGES, DIRECT OR INDIRECT, ARISING FROM OR RELATED TO USERS
FALURE TO FOLLOW THE INSTRUCTIONS SET FORTH ABOVE.

M0046_H

6

G10VPL

GUS Disinfection Soak Station™

Summary of Contents for GUS G10VPL

Page 1: ...741 WWW CIVCO COM Copyright 2018 All Rights Reserved GUS is a registered trademark of CIVCO Medical Solutions All other trademarks are property of their respective owners Printed in the USA G10VPL Wal...

Page 2: ...Mounting Instructions 4 Using High Level Disinfectants with GUS 6 Step 1 Probe Soaking Instructions 6 Step 2 Filling the Containers 7 Step 3 Ready for Use 7 Repair and Modification 8 Filter Life 9 Fi...

Page 3: ...ed Safety Precautions Disconnect power before replacing filter or when servicing the unit Refer to the MSDS of the high level disinfectant that you are using for additional safety precautions Environm...

Page 4: ...inst unsafe practices or potential equipment damage NOTE Indicates additional helpful information Quick Troubleshoot If the GUS system is not running check the following 1 Ensure that the unit is plug...

Page 5: ...ns This MUST be placed above the workstation l One Bottom support bracket l One filter installed in unit Wall Mounting Instructions 1 You will need the following items before attempting to install thi...

Page 6: ...p 5 9 Place the unit back over the two top screws Screw the two remaining Philip screws into the bottom bracket and tighten Tighten the top screws CAUTION Do not over tighten the screws 10 For masonry...

Page 7: ...ble as follows If your probe is NOT water resistant Suspend the probe into an empty container so that the top of the handle is level with the top of the container Depress blue tab and place probe cabl...

Page 8: ...as there are chemicals inside the containers 5 Fill the remaining container with clean rinse water for the preliminary rinse You MUST have one container for rinse water This ensures that when the pro...

Page 9: ...complete push back on the blue tab and lift the probe up shake it gently to remove large drops of disinfectant Then place it directly into the rinse container while keeping your thumb on the red mark...

Page 10: ...ting to the rear of the machine 6 Put the used filter into a plastic bag and dispose of it with other non hazardous waste This does not need to be red tagged 7 Replace the top plate and tighten the ca...

Page 11: ...spills CIVCO Medical Solutions manufactures Glute Out a glycine based neutralizer part 610 1099 which comes in a convenient single use bottle Glute Out is specifically for use with glutaraldehyde and...

Page 12: ...BILITY AND OF FITNESS FOR A PARTICULAR PURPOSE ARISING FROM A COURSE OF DEALING OR USAGE ARE SPECIFICALLY EXCLUDED This warranty does not cover conditions and damages resulting from any of the followi...

Reviews: