Citizon CKG Operating Manua Download Page 6

CKG counting scales operation manual                                                                           

REV C1, June 2012 

                                                                              CKG counting scales operation manual 

5.3 Automatic Part Weight Updates 

The scales will automatically update the unit weight when a sample 

equal to less than the sample already on the platform is added. A beep 

will be heard when the value is updated. It is wise to check the quantity is 

correct when the unit weight has been updated automatically. 

This feature is turned off as soon as the number of items added 

exceeds the count used as a sample. 

 

6. WEIGHING AND COUNTING CHECKING ALARM 

6.1 Set Hi/low Limit 

Press 

Pst

 key, display show 

Hi cnt

, key in high limit for counting 

alarm, press 

TARE

 to confirm, then display show 

lo cnt

, key in low limit 

for counting alarm, press 

TARE

 to confirm, then display show 

hi net

key in high limit for weighing, press 

TARE

 to confirm, then display   

 

show 

lo net

, key in low limit for weighing, press 

TARE

 to confirm, back 

to normal weighing mode. 

6.2 Checking Alarm 

At first, you need set checking mode, hold 

CE

 key, you will show 

check net

 (weighing checking alarm), 

check pcs

 (counting checking 

alarm) or 

check off

 (don’t use alarm function). 

You can set beep when high (when weight/quantity> high limit ,beep 

on),low(when weight/quantity<low, beep on), OK(when weight/quantity in 

high-low range, beep on),NG(when weight/quantity out of high-low range, 

beep on), about how to set the beep mode, see detail in technical 

manual.  

 

7. ACCUMULATED OPERATION 

7.1 Manually Accumulated Total 

The values (weight and count) shown on the display can be added 

 

-6- 

to the values in the accumulator by pressing the 

M+ 

key. The "Weight " 

display will show the total weight, the "Count" display will show the total 

accumulated count and the "Unit Weight" display shows the number of 

times items have been added to the accumulation memory. The values 

will be displayed for 2 seconds before returning to normal. 

The scale must return to zero or a negative number before another 

sample can be added to the memory.   

More product can then be added and the 

M+

 key pressed again. 

This can continue for up to 99 entries, or until the capacity weight display 

is exceeded. 

To observe the total stored press the 

MR 

key.  The totals will be 

displayed for 2 seconds. 

To clear the memory press

 MR 

to recall the totals from memory and 

then press the 

CE

 key to clear all values from memory. 

 

7.2 Automatic Accumulated Totals 

The scale can be set to automatically accumulate totals when a 

weight is placed on the scale. This eliminates the need to press the 

M+ 

key to store values into memory. However the 

M+

 key is still active and 

can be pressed to store the values immediately.    In this case the values 

will not be stored when the scale returns to zero. 

 

8. CALIBRATION 

Press 

TARE

 key during selfchecking, to enter the parameter menus, 

the default password is 0000. 

When the parameter menu show "

F1 CAL

" press the 

TARE

 key.     

The display will then show "unLoAd" to request all weight be removed 

from the platform.    

Press the 

TARE

 key to set the zero point. 

Key in the calibration weight you want to use, then put this weight on 

the platform and then press the 

TARE

 key. The scale should be   

-7- 