Make sure to check the following before sending your blood pressure monitor away 
to be tested or repaired.

Troubleshooting (1)

Problem

Possible Cause

Solution

LCD Display 

shows abnormal 

result

The cuff position was not correct or 
it was not properly tightened

Apply the cuff correctly and 
try again.

Body posture was not correct 
during testing

Review the “BODY POSTURE 
DURING MEASUREMENT” 
sections of the instructions and 
re-test.

Speaking, arm or body 
movement, angry, excited or 
nervous during testing

Re-test when calm and without 
speaking or moving during 
the test.

Irregular heartbeat
(arrhythmia)

It is inappropriate for people 
with serious arrhythmia 
to use this Electronic 
Sphygmomanometer.

Troubleshooting (2)

Problem

Possible Cause

Solution

LCD shows low 

battery symbol 

Low Battery

Change the batteries. 

LCD shows “Er 0”

Pressure system is unstable 

before measurement

Don’t move and try again.

LCD shows “Er 1” Fail to detect systolic pressure

LCD shows “Er 2” Fail to detect diastolic pressure

LCD shows “Er 3”

Pneumatic system blocked or 

cuff is too tight during inflation

Apply the cuff correctly and 

try again.

LCD shows “Er 4”

Pneumatic system leakage or 

cuff is too loose during inflation

LCD shows “Er 5”

Cuff pressure above 

300 mmHg

Measure again after five 

minutes. If the monitor is still 

abnormal, please contact the 

local distributor or the factory.

LCD shows “Er 6”

More than 3 minutes with cuff 

pressure above 15 mmHg

LCD shows “Er 7”

EEPROM accessing error

LCD shows “Er 8” Device parameter checking error

LCD shows “Er A”

Pressure sensor parameter 

error

No response when 

you press button 

or load battery.

Incorrect operation or strong 

electromagnetic interference

Take out batteries for five 

minutes, and then reinstall all 

batteries.

*  If you want to dispose this product, do not mix with general household waste. There is a 

separate collection systems for used electronics products in accordance with legislation.

BEFORE REQUESTING REPAIRS OR TESTING

00̲CHU304̲All.indb   Eng-15

2015/06/02   11:00:53

Summary of Contents for CHU304

Page 1: ...Espa ol English INSTRUCTION MANUAL FOR DIGITAL BLOOD PRESSURE MONITOR REF CHU304 CHU305 1501 English Espa ol Fran ais Py Please read all of the information in the instruction manual before operating...

Page 2: ...ing record of your blood pressure variations A complete picture can only be obtained from a large number of readings To obtain correct blood pressure measurements 1 Take five or six deep breaths and t...

Page 3: ...fications under the WHO standards Remarks The graph is not exact but may be used as a guide in understanding non invasive blood pressure measurements The device is only intended for use by adults Defi...

Page 4: ...ill be taken or who suffers from a frequent irregular heartbeat Consult your physician for advice on whether to use the monitor 9 DO NOT wrap the cuff around an injured arm 10 DO NOT wrap the cuff aro...

Page 5: ...as this could damage the monitor 8 To clean the cuff wipe it with a moist cloth Avoid hard rubbing as this will cause air leakages Take care also not to get water into the air hose 9 Clean the cuff a...

Page 6: ...isposal service or the point of sale where you purchased the items Information on Disposal in other Countries outside the European Union This symbol is only valid in the European Union If you wish to...

Page 7: ...hose socket START STOP Switch Stand by See page 7 for the Battery compartment See page 11 for the Irregular Heartbeat IHB Indicator Display Systolic blood pressure display Ready to inflate indicator...

Page 8: ...he compartment Battery cover Battery compartment 4 AA batteries 3 Close the battery cover When the icon is displayed or nothing appears in the display replace the old batteries with four new ones All...

Page 9: ...ress the START STOP switch to confirm the setting The month is set and the day indicator blinks You can fast forward the numbers in display by pressing and holding the MEMORY switch 4 Press the MEMORY...

Page 10: ...palm side Metal hoop 1 Unroll the cuff and thread the end through the metal hoop so that the side with the hook and loop fastener is on the outside The position of the artery 3 Adjust the position of...

Page 11: ...If the previous user had a high pressure setting that high pressure will automatically be used for the current measurement If the pressurization value exceeds 260 mmHg if pressurization feels abnormal...

Page 12: ...ected during measurement the Irregular Heartbeat IHB Indicator is displayed after measurement is completed There are 2 conditions under which the signal of IHB will be displayed 1 The coefficient of v...

Page 13: ...ur hand slightly on the table so that your palm is facing up and your fingers are relaxed 4 Do not move your body or talk while taking the measurement 99 measurement results can be stored in Memory Th...

Page 14: ...e latest data Data number 1 2 3 98 99 The latest data The earliest data HOW TO DELETE THE DATA 1 When any result except average reading of the last three measurements is displaying press and hold the...

Page 15: ...e ambient temperature or your mental state When you have a physician or nurse measure your blood pressure the pressure value is likely to be 10 to 20 mmHg higher than usual due to anxiety and or stres...

Page 16: ...measurement Don t move and try again LCD shows Er 1 Fail to detect systolic pressure LCD shows Er 2 Fail to detect diastolic pressure LCD shows Er 3 Pneumatic system blocked or cuff is too tight duri...

Page 17: ...idity 90 RH Electric shock protection Internal power unit Degree of Protection Type BF applied part Mode of operation Continuous operation Memory 1 99 readings Average of last 3 readings Service Life...

Page 18: ...oltage uctuations icker emissions IEC 61000 3 3 N A Guidance and manufacturer s declaration electromagnetic immunity The CHU304 CH305 is intended for use in the electromagnetic environment speci ed be...

Page 19: ...due to xed RF transmitters an electromagnetic site survey should be considered If the measured eld strength in the location in which the CHU304 CH305 is used exceeds the applicable RF compliance leve...

Page 20: ...CITIZEN is a registered trademark of Citizen Holdings Co Japan Design and Specification are subject to change without notice...

Reviews: