DISPLAY MARK
CONDITION/CAUSE
CORRECTIVE ACTION
Appears when the wrist cuff
is wrapped too loose.
Re-wrap the wrist cuff snugly.
Take another reading, making
sure you remain still.
(
See page 14
“
HOW TO
WRAP THE WRIST CUFF
”
.)
Appears when your wrist
moved during the measure-
ment.; or
Appears when the cuff is
not wrapped correctly.
Re-wrap the wrist cuff corre ctly.
Take another reading, making sure
you rem ain perfec tly stil l.(
See pag e 14
“
HO W TO WRA P
THE WRIST CUFF
”
.)
Contact the store where it was
purchased or your local service
center. (
See page 30
“
BEFORE
ASKING
FOR
REPAIR/INSPECTION
”
.)
Re-w rap the wrist cuff pro perly.
Take deep breaths to relax, and
measure again.
(
See page 13
“
POINTS FOR
ACCURATE MEASUREMENT
”
.)
Replace two alkaline batteries
with new ones.
(
See page 9
“
HOW TO
UNLOAD BATTERIES
”
.)
Cont act the st or e wher e it was
purcha sed or your local servi ce
cen ter. (
See pa ge 30
“
BEFORE
ASKI NG
FOR
REPAIR/INSPECTION
”
.)
Appears when the blood
pressure could not be
measured accurately.
Appears when the pulse
rate falls outside the
measuring range (below 39
beats/min. or above 181
beats/min.)
Appears when the batteries
have run out.
The unit does not function
properly.