After cleaning the cuff with a neutral detergent, lightly rinse it with
water and let it dry in the air. Hard rubbing or washing in a washing
machine will cause air leakage. Also, be careful not to get water into
the air hose.
SPECIFICATIONS
CLEANING METHOD OF CUFF
Measuring method:
Oscillometric system
Indication:
Digital display
Measuring range:
Pressure; 0
⬃
280mmHg, Pulse; 40
⬃
200 beat/min.
Accuracy:
Pressure;
±
3mmHg, Pulse;
±
5%
Inflation:
Automatic inflation by internal pump
Deflation:
Automatic constant speed deflation system
Exhaust:
Quick exhaust by electromagnetic valve
Memory function:
One data
Power source:
2 pcs “AAA” (LR03) alkaline batteries
Battery life:
Approx. 250 times/LR03 (once a day, 180mmHg, room temp. 22
°
C)
Rated voltage:
3V DC
Working temp. & humidity:
+10
°
C
⬃
+40
°
C, 30% to 85%RH
Storage temp. & humidity:
–20
°
C
⬃
+60
°
C, 10% to 95%RH
Left wrist circumference:
13.5cm
⬃
19.5cm
Outside dimensions:
74mm(W)
×
31mm(D)
×
72mm(H)
Weight:
Approx. 140g (including batteries)
Electric shock protection:
Type B equipment
Accessories:
2 pcs “AAA” (LR03) alkaline batteries, Instruction manual, Storage case
*Specifications and designs may be changed without notice in the event of improvements being made.
-
Eng 2
-