Resistance
Range

Resolution

Accuracy

200.0Ω

0.1Ω

±1.0% of reading ± 4 digits

2.000kΩ

±1.0% of reading ± 2 digits

20.00kΩ

10Ω

±1.2% of reading ± 2 digits

200.0kΩ

100Ω

2.000MΩ

1kΩ

20.00MΩ

10kΩ

±2.0% of reading ± 5 digits

200.0MΩ

10kΩ

±5.0% of reading ± 10 digits

Capacitance
Range

Resolution

Accuracy

2.000nF

1pF

±4.0% of reading ± 20digits

20.00nF

10pF

±4.0% of reading ± 3 digits

200.0nF

0.1nF

±4.0% of reading ± 3 digits

2.000uF

1nF

200.0uF

0.1uF

≤20uF

±4.0% of reading ± 15 

digits

≥21uF

unspecified

Input Protection: 250V dc or 250V ac rms.

Diode Test

Test current: 1mA typical

Open circuit voltage: 2.8V dc typical
Overload protection: 250V dc or ac rms.

Audible continuity

Audible threshold: Less than 50Ω Test current: <0.3mA
Overload protection: 250V dc or ac rms.

Summary of Contents for DMR-6600

Page 1: ...USER S MANUAL Digital Multimeter DMR 6600 CIRCUIT TEST ELECTRONICS www circuittest com...

Page 2: ...situation which if not avoided could result in death or serious injury This CAUTION symbol indicates a potentially hazardous situ ation which if not avoided may result damage to the product This symb...

Page 3: ...ly When using the test probes always keep fingers behind the finger guards Never touch the exposed probe tip Always consider circuits to be energized Never assume any equipment to be de energized Use...

Page 4: ...itance Continuity Diode 250V DC AC rms 2 USE EXTREME CAUTION when working with high voltages 3 NEVER connect the meter leads across a voltage source while the function switch is in the current resista...

Page 5: ...switch 4 20A positive input jack 5 A mA input jack 6 Backlight button 7 Positive input jack 8 COM negative input jack Continuity Diode Ohms Volts DC Volts AC Amps DC Amps AC F Farads micro farads m m...

Page 6: ...d nominal Operating Temperature 0 C to 50 C 32 F to 122 F at 70 relative humidity Storage temperature 20 C to 60 C 4 F to 140 F at 80 relative humidity Fuse 200mA 250V 5 x 20mm Fast blow 20A 250V 6 35...

Page 7: ...1V 1 2 of reading 5 digits Overload Protection 0 2A 250V and 20A 250V Fuse Note Accuracy is given at 18 C to 28 C 65 F to 83 F 70 RH AC Current Range Resolution Accuracy 2 000mA 1uA 1 2 of reading 3 d...

Page 8: ...ution Accuracy 2 000nF 1pF 4 0 of reading 20digits 20 00nF 10pF 4 0 of reading 3 digits 200 0nF 0 1nF 4 0 of reading 3 digits 2 000uF 1nF 200 0uF 0 1uF 20uF 4 0 of reading 15 digits 21uF unspecified I...

Page 9: ...e battery life of the meter the backlight will automatically turn off in approximate 30 seconds DC Voltage Measurements CAUTION Do not measure DC voltages if a motor on the circuit is being switched O...

Page 10: ...uch the black test probe tip to the negative side of the circuit and the red test probe tip to the positive side of the circuit 4 Read the voltage in the display The display will indicate the proper d...

Page 11: ...Touch the black test probe tip to the negative side of the circuit Touch the red test probe tip to the positive side of the circuit 6 Apply power to the circuit 7 Read the current in the display 8 For...

Page 12: ...acitors before taking any resistance measurements Remove the batteries and unplug the line cords 1 Set the function switch to the position 2 Insert the black test lead banana plug into the negative CO...

Page 13: ...lead banana plug into the negative jack COM and the red test lead banana plug into the positive jack 3 Touch the test leads to the capacitor to be tested 3 Touch the test probe tips to the diode or se...

Page 14: ...o not operate your meter until the battery door is in place and fastened securely Replacing the Fuses WARNING To avoid electric shock disconnect the test leads from any source of voltage before removi...

Page 15: ...period of 2 years from the date of purchase Any product which has been subjected to misuse or accidental damage is excluded from the warranty Except as stated above Circuit Test Electronics makes no p...

Reviews: