14. Maintain final hypothermia temperature by resetting the Cool set point
temperature to 10°C and turning off the pump when you reach the
hypothermia temperature desired. If the temperature should tend to rise,
reactivate the pump allowing the cool water to circulate through the heat
exchanger and as a result bring the blood temperature down again.
15. Preset your hot water temperature to the desired initial re-warming
temperature.
16. The units will re-warm by activating the "HEAT" switch. The Hemotherm will
receive the maximum wattage of heat to raise the hot water fluid temperature
to the set point temperature. At this point, if desired, connect your Hyper-
Hypothermia blanket to the 1/8" quick-connect fittings of the HEMOTHERM
unit to maximize re-warming efficiency.
To discontinue operation, simply turn off the power switch and disconnect all lines
leading to the heat exchanger and/or blanket from the HEMOTHERM unit.
1.9
REMOTE CONTROL OPTION
All HEMOTHERM units are equipped with a remote control port. The remote control port
connection is located on the right side of the unit, near the power switch/circuit breaker. If
the remote control option is purchased, the “assembly" will include a mating 25 foot cable,
the remote control unit, and a mounting clamp.
To install the remote control option, first turn the HEMOTHERM unit power switch off.
Mount the remote control in a convenient place utilizing the universal mounting clamp
provided. Connect the cable between the two devices and secure in place by tightening
the screws attached to the cable housing. The remote control will operate the same as
the key pad located on the HEMOTHERM main unit.
NOTE
If you happen to overshoot your desired "HEAT" water temperature, you may push the
"COOL" switch for five to ten seconds allowing the water in the water lines and heat
exchanger to empty into the "COOL" water reservoir. Switching back to the "HEAT"
mode will allow cool water in the water lines and heat exchanger to be mixed into the
"HEAT" water, in turn cooling the temperature of that water a few degrees (1°- 4°C)
depending on the difference in temperature between the two reservoirs. The same can
be done if you overshoot your desired cool water temperature.
NOTE
The Model 400CE will only work with the Model 414CE Remote control and cables and