9
User Menu Parameters
To Lock Screen on/off
When main screen is active,
Press
H+D
buttons together / 3s
to lock and unlock screen
on/oFF
)
T
H
U
D
loc
Alarm delay
dlY
%
T
H
U
D
Alr
When main screen is active,
keep pressed
U
button / 3s
Voice alarm will be delayed
15 minutes.
When main screen is active,
Press
U
button and release to check
remaining time for next cooling
Note:
remaining time to next cooling will appear
if periodic cooling function on
( pls check APC menu. (
See page 17
)
Than screen turns back to main screen.
Remaining time for next cooling
11.27
%
T
H
U
D
clt
50
%
T
H
U
D
37.7
50
%
T
H
U
D
37.7
10
%
T
H
U
D
hhı
-10
%
T
H
U
D
hlo
0,8
%
T
H
U
D
-2.0
%
T
H
U
D
-Press T + H together / 3s to enter the menu
-Use T /H buttons to pass next/before parameter.
-Use U/D buttons to change parameter values.
hhi: high humidity alarm.
Started :
set value +10
Delay time:
20 minutes
Notification:
Alr 5
Voice alarm :
continuous
hlo: low humidity alarm.
Started :
set value -10
Delay time:
20 minutes
Notification:
Alr 6
Voice alarm : intermittently
(f)thi: high temperature alarm.
Started :
set value + 0,8
Delay time:
2 minutes
Notification:
Alr 1/3
Voice alarm :
continuous
(f)thi
(f)tlo
0.0
%
T
H
U
D
hco
0.0
%
T
H
U
D
(f)tco
PD SERIES
(f)tlo: low temperature alarm.
Started :
set value - 2.0
Delay time:
30 minutes
Notification:
Alr 2/4
Voice alarm : intermittently
tco: temperature calibration value.
CAUTION!
tCo is very critical parameter and
changing this can results with chick looses. pls use
qualified
and
calibrated
glass
or
electronic
thermometers for calibration purpose. And always
check temperature values periodically.
hco: humidity calibration value.
CAUTION!
hCo is very critical parameter and
changing this can results with chick looses. pls use
qualified
and
calibrated
glass
or
electronic
thermometers for calibration purpose. And always
check humidity values periodically.
-Press T + H together / 3s to save changes.
donE
%
T
H
U
D
SAVE