Freeze-dryer Epsilon 1-4 LSCplus
Freeze-dryer Epsilon 2-4 LSCplus
5 Set-up and connection
Version 04/2014, Rev. 2.13 of 01/12/2021
• sb
47 / 148
Translation of the original operating manual
• Do not set up the vacuum pump directly next to the heat exchanger
(condenser).
• In the case of water-cooled systems, ensure that the water circuit
provides a sufficient amount of cooling water.
• Avoid direct sunlight (UV radiation).
Pos : 132 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 133 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 134 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/050 Aufst ellung und Ansc hluss/ 050 Aufst ellung und Ans chl uss Epsilon/ 050-0020 Ver- und Ents orgungs ansc hlüss e Epsilon @ 48\m od_1490088361900_68.docx @ 492302 @ 2 @ 1
5.2
Supply and disposal connections
Pos : 135 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/050 Aufst ellung und Ansc hluss/ 050 Aufst ellung und Ans chl uss Epsilon/ 050-0020- 0010 Energiev ers orgung Epsilon --------------------------------------- @ 48\mod_1490087247571_68. docx @ 492266 @ 3 @ 1
→
5.2.1
Power supply
Pos : 136 /200 Chris t/361 GT-BA Labor-Pil ot ( PROJ EKTE)/ Epsilon 4er/ Epsilon 1-4_2- 4 LSC plus Standar d/ 050 Aufst ellung und Ansc hl uss/ 050- 0020- 0010-0010 Ans chl uss art neu E1- 4_2- 4 @ 48\m od_1490346443622_68.docx @ 493183 @ 4 @ 1
→
5.2.1.1
Connection
DANGER
The operating voltage on the name plate must correspond to the local
supply voltage
Christ freeze-dryers are units of safety class I. Epsilon 1-4 LSCplus and
Epsilon 2-4 LSCplus units have a three-wire power cord with a
CEE 7/4 plug (type F) (see chapter 10 - "Technical data").
Pos : 137 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 138 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/050 Aufst ellung und Ansc hluss/ 050 Aufst ellung und Ans chl uss Epsilon/ 050-0020- 0010-0020 Sic herungen baus eits Epsilon @ 25\m od_1404984623550_68.docx @ 185665 @ 4 @ 1
→
5.2.1.2
Customer-provided fuses
The freeze-dryer must be protected by way of a customer-provided mains
fuse with the correct rating.
Pos : 139 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 140 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/050 Aufst ellung und Ansc hluss/ 050 Aufst ellu ng und Ans chl uss Epsilon/ 050-0020- 0020 Belüftung Epsil on 1 Ansc hlus s @ 48\m od_1490090398173_68.docx @ 492320 @ 3 @ 1
→