Freeze-dryer Alpha 1-2 LSCbasic
6 Operation
70 / 111
Version 07/2022, Rev. 1.0 of 19/07/2022
• sb-ja
Translation of the original operating manual
NOTE
If the value "∞" (infinite) is selected for a process phase, the next phase
must be started manually via the button "Operating mode: select/start".
The set values of the active phase can be changed during the process run.
In this case, the control system adapts the freeze-dryer to the new set
values as quickly as possible.
After the completion of a phase, the freeze-dryer switches to the next
phase without switching to standby. The transition from "Freezing" to
"Warm-up" takes place automatically. After the completion of the "Warm-
up" phase, a dialogue box opens:
Fig. 53: Dialogue box after the completion of the warm-up phase
The freeze-dryer will remain in the "Warm-up" phase until a confirmation is
issued.
Normally, the transition from "Main drying" to "Final drying" takes place
automatically.
After the completion of the "Final drying" phase, there will be another
enquiry with which the freeze-drying process will be completed. The freeze-
dryer remains in the "Run" mode until the enquiry is confirmed.
The process can be stopped any time by way of the "Stop" button. In this
case, the freeze-dryer will be switched to standby.
6.5.2.1
Entering set values
The system has stored set values for every phase, and for every value
there are pre-defined value ranges that can be determined in the various
dialogue boxes by pressing the buttons "min" and "max".
In order to protect the product, a
→
safety pressure
value can be entered in
every drying section.