Freeze-dryer Alpha 1-2 LSCbasic
5 Set-up and connection
Version 07/2022, Rev. 1.0 of 19/07/2022
• sb-ja
43 / 111
Translation of the original operating manual
5. 4. 6
Vak uumpumpe
→
5.5
Vacuum pumps
To ensure its intended use, the freeze-dryer Alpha 1-2 LSCbasic must be
used with one of the following vacuum pumps:
Oil-sealed vacuum pumps:
• Rotary vane pump Vacuubrand RZ2.5
• Rotary vane pump Edwards RV3
5. 4. 6
Vak uumpumpe
→
5.5.1
Oil-sealed vacuum pump and exhaust filter (oil mist separator)
NOTE
It is absolutely essential to refer to the separate instruction manual of the
vacuum pump and exhaust filter (if applicable)!
The vacuum pump must be connected to the vacuum connection of the unit
and to the electrical socket at the back of the unit, which is marked
accordingly (see chapter 2.1.1 - "Functional and operating elements").
NOTE
The vacuum pump is supplied with power by the unit, but the maximum
current for the vacuum pump is limited. It is absolutely essential to refer to
the label of the electrical outlet for the vacuum pump (see the following
picture)!
If the current requirement of the vacuum pump is higher than the value
that is stated on the label, the pump must be supplied separately via an
on-site power socket. Alternatively, it can be connected to the freeze-dryer
via an interface box (see chapter 5.6 - "Use of an interface box").
1
Label indicating the
maximum current
Fig. 13: Indication of the maximum current for the vacuum pump (example)