3.5 End of test and cleaning of the Rheo F4
The Rheo F4 will bring up a warning screen to inform you that the
test is finished. You can save the test by clicking the button:
[Save]
After the test and before conducting a new test, you must clean the
Rheo F4.
The bowl, the perforated basket and the weights must be kept
clean at all times. After each test, clean the basket and the weights
using a damp sponge. The bowl must be cleaned using a damp
cloth to prevent water entering the pressure tap at the bottom of
the bowl. In particular thoroughly clean the bowl rubber seal and
the lid surface.
After each test, use a dry cloth to wipe all potentially
damp parts that could prevent the device from operating
properly. Check that the soda lime in the absorption
cartridge is not saturated (blue color). For more
information, refer to the maintenance manual. When the
device is not in use, never hermetically close the bowl. To
avoid condensation, let the lid rest on the studs.