Calibrations
❏
This menu allows to modify some of the calibrations
(unapproved) and to enter new calibrations.
❏
Click on "Meas. and calibrations" and choose "Calibrations".
❏
Only unapproved calibrations ( "approved" box unticked) can be
modified.
A calibration is form:
Lin : rough measurement obtained with the linear calibration
❏
To modify a calibration, double-click on the box to be modified,
enter the modification and click on "PC => Aqua-TR" to update the
Aqua-TR memory.
❏
To insert a new calibration :
- Click on "Insertion", enter the name of the new calibration and click
on "Validate". The name of the new calibration must not have more
than 2 consecutive characters in common with an approved
calibration.
- Enter the parameters of the new calibration (A0, A1, A2 and A3),
the range of the calibration (Mmin, Mmax), the HLW ref. and the
HLW. Entering Mmin,Mmax, HLW ref. and
HLW is optional.
- Click on "PC => Aqua-TR" to transfer the new calibration.
Note:
For any modification on approved calibration, contact your Chopin
Technologies entitled repair organization.
H2O
A0 A1 Lin
A2 Lin
2
100
-----------
+
A3
Lin
3
10000
----------------
+
+
=
Summary of Contents for Aqua-TR
Page 2: ......
Page 4: ......
Page 5: ...Aqua TR 5 Mode d emploi 07 2015 F AQUA TR MODE D EMPLOI...
Page 8: ...Aqua TR 8 Mode d emploi 07 2015 F...
Page 47: ...Aqua TR 47 Mode d emploi 07 2015 F avec votre revendeur...
Page 50: ...Aqua TR 50 Mode d emploi 07 2015 F...
Page 51: ...Aqua TR 51 User s manual 07 2015 GB AQUA TR USER S MANUAL...
Page 54: ...Aqua TR 54 User s manual 07 2015 GB...
Page 95: ...Aqua TR 95 Modo de empleo 07 2015 E AQUA TR MODO DE EMPLEO...