INSTRUCTION  MANUAL

PULSE  OXIMETER

MD300I

MD300I

- 6 -

- 7 -

● OLED display screen with adjustable backlight.
● Up to 10 patients’ IDs and 72-hour records can be saved.
● Data transfer to PC by USB cable.
● Powered by two AA alkaline.
● Suitable for adult and pediatric patients.

CHAPTER 2  General Descriptions

2.1 Appearance

Fig.2-1 Appearance of the oximeter

Sensor Interface: This interface is designed for two purposes, one is to connect the SpO2 
probe for measurement, and the other is to connect the pulse oximeter to PC with USB cable 
for data transmission.
Function key - at the left of the oximeter, acts as a power on switch when the oximeter is 
power off. While the oximeter is power on, it acts as a function key.
Setting key - at the right of the oximeter, while the oximeter is power on, it acts as a setting 
key.
While there is no measurement or operation the oximeter will power off automatically in 30 
seconds.

2.2 Battery Installation

1. Slide the battery door cover along the arrow as shown in Fig.2-2.
2. Install two AA batteries into the battery compartment. Match the plus (+) and minus (-) 
signs in the compartment. If the polarities are not matched, damage may be caused to the 
oximeter.
3. Replace the battery cover.

 Warning!

Summary of Contents for MD300I

Page 1: ...PULSE OXIMETER INSTRUCTION MANUAL MD300I...

Page 2: ...to is protected by copyrights or patents and does not convey any license under the patent rights of our company nor the rights of others We do not assume any liability arising out of any infringements...

Page 3: ...follows A mathematical formula is established making use of Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive hemoglobin RHb and Oxyhemoglobin HbO2 in glow and near infra...

Page 4: ...accompanying documents This equipment should not be used adjacent to or stacked with other equipment This equipment is not intended for use during patient transport outside the healthcare facility Whe...

Page 5: ...Venous pulsations Placement of a sensor on an extremity with a blood pressure cuff arterial catheter or intravascular line The patient has hypotension severe vasoconstriction severe anemia or hypother...

Page 6: ...PC with USB cable for data transmission Function key at the left of the oximeter acts as a power on switch when the oximeter is power off While the oximeter is power on it acts as a function key Setti...

Page 7: ...sor cable upside when inserting the sensor 2 Do not insert the sensor forcefully CHAPTER 3 Take a Measurement Definition of key press There are three ways to press the key 1 Press press the key quickl...

Page 8: ...l warning SpO2 and PR If the measured SpO2 and or PR value is beyond the default limit the corresponding value will flash with audible sound By default SpO2 The upper limit 100 The lower limit 90 PR T...

Page 9: ...he Information Display mode you can set Brightness Patient ID Date and Time Current parameters and data will be displayed at the top right corner which is used to display the PR The various parameters...

Page 10: ...Definitions Err 1 program memory damaged Err 2 data memory damaged Err 3 sensor Red Emission Diode damaged Err 4 sensor Infrared ray Emission Diode damaged Err 5 sensor Infrared ray Receipt Diode dam...

Page 11: ...belong to consumption and beyond free change range Such damage of probe caused by mechanical force doesn t belong to free change range During measurement of SPO2 principle leads to measuring value di...

Page 12: ...ting Temperature 5 40 Storage Temperature 20 55 Operating Humidity 80 RH no condensation Storage Humidity 93 RH no condensation Classification per IEC60601 1 Classification according to IEC 60601 1 Ac...

Page 13: ...e model MD300I Pulse Oximeter uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment 5 RF e m...

Page 14: ...in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3Vrms 150kHz to 80MHz 3V m 80MHz to...

Page 15: ...r as recommended below according to the maximum output power of the communications equipment Rated maximum output of transmitter W Separation distance according to frequency of transmitter m 150KHz to...

Reviews: