6.4.4 Modulation degree programming
When the "dELy" is selected in the signal modulation mode, the first-order lag degree can be adjusted.
Further, when the “PEAk” is selected, the damping degree of signal after tracking the value can be
adjusted.
6.4.4-1 Modulation time constant programming
(
Effective when the “dELy” is selected in the signal modulation mode.
)
When the input signal is fluctuated, the measured value is displayed smoothly.
1) To program the modulation time constant, press
SEL
key in
the measurement mode screen until “
tAu
” appears in the sub
display as shown in the right figure.
2) By pressing
key, the most significant digit of the main
display will blink.
3) Use
△
or
▽
key to program the desired value.
4) To shift to the next lower digit, press
key.
The next lower digit will blink.
5) Repeat the above procedure of (3) and (4) up to the least
significant digit for programming (4 digits).
6) To activate this programming, press
ENT
key.
6.4.4-2 Damping degree selection
(Effective when the "PEAk" is selected in the signal modulation mode.)
1) To select the damping degree, press
SEL
key in the
measurement mode screen until “
dEc
” appears in the sub
display as shown in the right figure.
2) By pressing
key, “
0
” at the least significant digit in the
main display will blink.
3) Use
△
or
▽
key to select the desired damping degree
[0, 2, 5 or 10(ºC/second)].
4) To activate this selection, press
ENT
key.
•
By selecting
“dELy”
and programming to
“0.0”,
the measured value based on
the original signal is displayed.
Reference
・
The programmable range is
0.0
to
99.9
(second).
・
The default value is
0.0
second (original signal).
R e m a r k s
• The default value is
0
ºC (ºC/second).
R e m a r k s
Sub display
Main display
Tb
CONT
MEM
PEAK
AL
AH
C
F
ε
Sub display
AL
AH
C
F
ε
Tb
CONT
MEM
PEAK
Main display
Caution
Execute the programming of the modulation degree together with [6.4.3 Signal
modulation mode selection].