Chino IR-BAT Series Instructions Manual Download Page 6

INST No. INE-412-P7CE

1) Regularly check the

(2)

Lens for dust or dirt. If dirt is present, remove it with a

blower for camera lenses. If the dirt cannot be removed with the blower, wipe the

lens gently with cotton ball soaked in alcohol.

2) Check if the thermometer targets the measurement object.
3) Check if the "lead wires" is connected tight.

Please refer the table below if a trouble happens.

If this thermometer still does not work properly, contact CHINO's agent.

9.1 Setting

9.2 Connection

9.3 Power supply

9.4 Warm up

Problems

Cause

Solutions

Power supply lamp
does not light.

1) Power is not supplied.
2) Power voltage is low.

1) Check the connection of lead wires.
2) Check the power voltage.

Measured value is
incorrect.

1) Loosen connections of lead wires
2) Noises affect measurement.
3) Emissivity (

) setting digital

switch is incorrect.

4) A heat source affects measurement

5) The lens is dirty.
6) Un adequate target.

1) Tighten any loosen connections of lead

wires.

2) Move this thermometer from noise sources.
3) Set “Emissivity” correctly.
4) Insulate the measured object from the heat

sources.

5) Remove the dirt from the lens.
6) Adjust the target.

7. Check and maintenance

C au t i on

When wiping out the dirt of the lens, don't use organic solvents except

alcohol, water and detergent absolutely. If used, it will cause the trouble.

8. Trouble shooting

9. Cautions

C au t i on

1) Install this thermometer at places where the ambient temperature is 0 to 50

ºC

.

2) Avoid to sets this thermometer at places where there are dust or other airborne

particles.

3) Avoid to sets this thermometer at places with vibration or impact.

4) Avoid to sets this thermometer at places where there is dangerous gas such as

combustible explosion gas.

C au t i on

1) If the line voltage fluctuates over the range of ±10 % of the rated voltage,

you will have malfunction or trouble.

2) Avoid the use of the power source in which the spike -shape noise occurs

when it is turned on or off.

C au t i on

This thermometer works immediately when the power supply is turned on,

but it is necessary to warm up this thermometer for about 30 minutes to

have the accurate measurement.

C au t i on

Don't wire 3)-(5) cable near a noise occurrence source, relay drive lines or high
frequency lines

Avoid to bunch these lines with lines having noise or place them in

the same duct which the lines having noise are placed.

Summary of Contents for IR-BAT Series

Page 1: ... necessary for maintenance too Keep this manual with due care until this thermometer is discarded NOTE 1 The descriptions of this manual are subject to change without notice 2 If a question has arisen or if an omission was found in this manual please contact your nearest CHINO s sales agent 3 We are not responsible for any results by operation of this thermometer Thank you for your purchase of IR ...

Page 2: ...l be dangerous to causing the trouble Cautions on maintenance and check of the lens When wiping out the dirt of the lens don t use organics solvents except an alcohol water and the detergents absolutely If used it will cause the trouble Type Model Measuring range Response time Diameter Distance Measuring wavelength General General type IR BAT1A 0 to 300 ºC 100ms Ø40 500mm 8 to 14 m IR BAT1B 0 to 6...

Page 3: ...h part name and outside dimensions 6 7 Control panel 15 85 f 26 I N F R A R E D T H E R M O M E T E R I R B 21 1 55 32 22 2 3 4 5 1 25 1 13 10 11 12 4 M2 Screw for control panel cover 8 2 M4Depth 4 9 1 4w Depth 4 BA019T 004 IR BAT1A 0 300 03 BA044T 004 IR BAT1A 0 300 03 Temperature range Software version Model Serial No Serial No Label CE mark 1 V shaped groove 2 Lens 3 Sight line 4 3mm O tip 5 Ca...

Page 4: ...s and explosion gas Place this thermometer at the distance where the diameter of the measurement object is more than 1 5 times of the measurement diameter calculated by the following relation When the temperature of the measurement object is higher than the ambient temperature change the direction of this thermometer slightly to the direction having the maximum output On this direction you will ha...

Page 5: ... IR GBHA1 Indicator with power supply and IR GC1 Indicator with power supply and Model IR GBG2 Converter with reflection compensation refer to the instruction manual for each model Reference 6 Operation Caution If the temperature of the measurement object is unknown measure it with a thermocouple and adjust Emissivity setting digital switches Adjustments may not be possible in the following cases ...

Page 6: ...e heat sources 5 Remove the dirt from the lens 6 Adjust the target 7 Check and maintenance Caution When wiping out the dirt of the lens don t use organic solvents except alcohol water and detergent absolutely If used it will cause the trouble 8 Trouble shooting 9 Cautions Caution 1 Install this thermometer at places where the ambient temperature is 0 to 50ºC 2 Avoid to sets this thermometer at pla...

Page 7: ...0 Inner diameter Ø4 Connection cable inlet Purge air inlet Outer diameter Ø6 L metal fittings Ø 4 3 10 5 52 24 125 33 45 Ø 29 40 39 113 2 hole for mounting L angle After opening case fix it To the thermometer top 2 57 27 80 10 2 Protective case Thermometer by M4 screw Inner diameter Ø4 Purge air inlet Outer diameter Ø6 Ø 4 3 10 5 Connection cable inlet L metal fittings 2 hole for mounting L angle ...

Page 8: ...2000mm IR ZBRA Max length is 200m Out of CE Marking Case Aluminum die cast Waterproof JIS C 0920 Water drop protection II type IP X2 Mounting method By 2 M4 screws or on tripod Weight About 220g CE EN61326 1 Class A Conditions for CE conformance 1 The connection cable should be used indoors and its length should be up to 30m 2 One 1 set of DC power source unit should be individually connected to o...

Reviews: