Chino IR-BAT Series Instructions Manual Download Page 5

INST No. INE-412-P7CE

1

2

3

4

5

6

7 8 9

0

1

2

3

4

6

7 8 9

0

5

×1.0

×0.1

×0.01

Emissivity "

ε

" setting digital switches

In case of 1 to 0

In case of 0 to 1

I N F R A R E D T H E R M O M E T E R

I R- B

Recorder,

S+(White)

S- (Green)

P+(Red)

P- (Black)

E(Shield)

Controller

DC power

supply

12 to 24V DC

1) Turn on the power supply after making sure all connections are adequate.

(6) Power supply indicator lamp is lighted.

2) Remove the (7) Control panel cover for the control panel and adjust (11),(12),(13) Emissivity (

)

setting digital switches so that the analog output matches the temperature of the measurement object.

Marker

Cable color

Item

Remarks

S+

White

Analog output (+)

4 to 20mA DC

S-

Green

Analog output (-)

Load resistance Less than 280

P+

Red

Power supply (+)

12 to 24V DC (±10%) 60mA

P-

Black

Power supply (-)

E

Clear (Shield)

Ground (

E

)

Ground

5. Connection

C au t i on

Supply voltage: 12 to 24V DC (±10%), 60mA.
Don't short each lead wires. Lead wires not using should be taped or cut off.

Lead wires not using should be taped or cut off.
For

connection

with

Model

IR-GBHA1

(Indicator

with

power

supply),

and

IR-GC1(Indicator with power supply) and Model IR-GBG2(Converter with reflection
compensation), refer to the instruction manual for each model.

Reference

6. Operation

Cau tion

If the temperature of the measurement object is unknown, measure it with a thermocouple

and adjust Emissivity (”

) setting digital switches.

Adjustments may not be possible in the following cases:

1) When the reflectance ratio of the measurement object is high like as a metal gloss surface.
2) When there is a strong heat source near the measurement object.

Attention to the direction that turns it, in case of

setting the 1.0 column

The turn corner is about 250 degree

Attention not to turn too much, in case of using the

driver of accessory

Caution

Warnin g

For connection work, make sure to turn off the power supply for preventing an electric
shock accident.

For measuring temperature with the infrared radiation thermometer, it is necessary to set
the emissivity in accordance with the measurement object for accurate readings. Emissivity
varies depending on the material and surface condition of the measurement object, and you
need to set the emissivity in accordance with the measurement object.

Reference

Summary of Contents for IR-BAT Series

Page 1: ... necessary for maintenance too Keep this manual with due care until this thermometer is discarded NOTE 1 The descriptions of this manual are subject to change without notice 2 If a question has arisen or if an omission was found in this manual please contact your nearest CHINO s sales agent 3 We are not responsible for any results by operation of this thermometer Thank you for your purchase of IR ...

Page 2: ...l be dangerous to causing the trouble Cautions on maintenance and check of the lens When wiping out the dirt of the lens don t use organics solvents except an alcohol water and the detergents absolutely If used it will cause the trouble Type Model Measuring range Response time Diameter Distance Measuring wavelength General General type IR BAT1A 0 to 300 ºC 100ms Ø40 500mm 8 to 14 m IR BAT1B 0 to 6...

Page 3: ...h part name and outside dimensions 6 7 Control panel 15 85 f 26 I N F R A R E D T H E R M O M E T E R I R B 21 1 55 32 22 2 3 4 5 1 25 1 13 10 11 12 4 M2 Screw for control panel cover 8 2 M4Depth 4 9 1 4w Depth 4 BA019T 004 IR BAT1A 0 300 03 BA044T 004 IR BAT1A 0 300 03 Temperature range Software version Model Serial No Serial No Label CE mark 1 V shaped groove 2 Lens 3 Sight line 4 3mm O tip 5 Ca...

Page 4: ...s and explosion gas Place this thermometer at the distance where the diameter of the measurement object is more than 1 5 times of the measurement diameter calculated by the following relation When the temperature of the measurement object is higher than the ambient temperature change the direction of this thermometer slightly to the direction having the maximum output On this direction you will ha...

Page 5: ... IR GBHA1 Indicator with power supply and IR GC1 Indicator with power supply and Model IR GBG2 Converter with reflection compensation refer to the instruction manual for each model Reference 6 Operation Caution If the temperature of the measurement object is unknown measure it with a thermocouple and adjust Emissivity setting digital switches Adjustments may not be possible in the following cases ...

Page 6: ...e heat sources 5 Remove the dirt from the lens 6 Adjust the target 7 Check and maintenance Caution When wiping out the dirt of the lens don t use organic solvents except alcohol water and detergent absolutely If used it will cause the trouble 8 Trouble shooting 9 Cautions Caution 1 Install this thermometer at places where the ambient temperature is 0 to 50ºC 2 Avoid to sets this thermometer at pla...

Page 7: ...0 Inner diameter Ø4 Connection cable inlet Purge air inlet Outer diameter Ø6 L metal fittings Ø 4 3 10 5 52 24 125 33 45 Ø 29 40 39 113 2 hole for mounting L angle After opening case fix it To the thermometer top 2 57 27 80 10 2 Protective case Thermometer by M4 screw Inner diameter Ø4 Purge air inlet Outer diameter Ø6 Ø 4 3 10 5 Connection cable inlet L metal fittings 2 hole for mounting L angle ...

Page 8: ...2000mm IR ZBRA Max length is 200m Out of CE Marking Case Aluminum die cast Waterproof JIS C 0920 Water drop protection II type IP X2 Mounting method By 2 M4 screws or on tripod Weight About 220g CE EN61326 1 Class A Conditions for CE conformance 1 The connection cable should be used indoors and its length should be up to 30m 2 One 1 set of DC power source unit should be individually connected to o...

Reviews: