INST No. INE-412-P7CE
1
2
3
4
5
6
7 8 9
0
1
2
3
4
6
7 8 9
0
5
×1.0
×0.1
×0.01
Emissivity "
ε
" setting digital switches
In case of 1 to 0
In case of 0 to 1
I N F R A R E D T H E R M O M E T E R
I R- B
Recorder,
+
-
+
-
E
S+(White)
S- (Green)
P+(Red)
P- (Black)
E(Shield)
Controller
DC power
supply
12 to 24V DC
1) Turn on the power supply after making sure all connections are adequate.
(6) Power supply indicator lamp is lighted.
2) Remove the (7) Control panel cover for the control panel and adjust (11),(12),(13) Emissivity (
”
”
)
setting digital switches so that the analog output matches the temperature of the measurement object.
Marker
Cable color
Item
Remarks
S+
White
Analog output (+)
4 to 20mA DC
S-
Green
Analog output (-)
Load resistance Less than 280
P+
Red
Power supply (+)
12 to 24V DC (±10%) 60mA
P-
Black
Power supply (-)
E
Clear (Shield)
Ground (
E
)
Ground
5. Connection
C au t i on
Supply voltage: 12 to 24V DC (±10%), 60mA.
Don't short each lead wires. Lead wires not using should be taped or cut off.
Lead wires not using should be taped or cut off.
For
connection
with
Model
IR-GBHA1
(Indicator
with
power
supply),
and
IR-GC1(Indicator with power supply) and Model IR-GBG2(Converter with reflection
compensation), refer to the instruction manual for each model.
Reference
6. Operation
Cau tion
・
If the temperature of the measurement object is unknown, measure it with a thermocouple
and adjust Emissivity (”
”
) setting digital switches.
・
Adjustments may not be possible in the following cases:
1) When the reflectance ratio of the measurement object is high like as a metal gloss surface.
2) When there is a strong heat source near the measurement object.
・
Attention to the direction that turns it, in case of
setting the 1.0 column
The turn corner is about 250 degree
・
Attention not to turn too much, in case of using the
driver of accessory
!
Caution
Warnin g
For connection work, make sure to turn off the power supply for preventing an electric
shock accident.
!
For measuring temperature with the infrared radiation thermometer, it is necessary to set
the emissivity in accordance with the measurement object for accurate readings. Emissivity
varies depending on the material and surface condition of the measurement object, and you
need to set the emissivity in accordance with the measurement object.