![Chiappa Arredamenti 120765 SLB RD Instructions For Use Manual Download Page 55](http://html1.mh-extra.com/html/chiappa-arredamenti/120765-slb-rd/120765-slb-rd_instructions-for-use-manual_2597879055.webp)
9.
Electromagnetic Compatibility (EMC)
EMC Test report
Conducted emission in the frequency range 150kHz-30MHz
The graphs 1, 2, 3, 4 indicates that the equipment considered in working cycle meets the
EN 60601-1-2 requirement.
Radiated emission in the frequency range 30MHz-1000MHz
The graphs 5, 6, 7, 8 indicate that the equipment consider in working cycle meet the EN
60601-1-2 requirement.
Harmonic measurement
The result indicate that the equipment consider in working cycle meet the EN 60601-1-2
requirement.
Flicker measurement
The result indicate that the equipment consider in working cycle meet the EN 60601-1-2
requirement.
Fast transients BURST
The result indicate that the equipment consider in working cycle meets the EN 60601-1-2
requirements.
Electro-static discharge immunity test
The result indicate that the equipment consider in working cycle meets the EN 60601-1-2
requirements.
RF conducted immunity test
The result indicate that the equipment consider in working cycle meets the EN 60601-1-2
requirements.
IFU-120765-SLB-RD-Chiappa-Arredamenti-11/2017#ed04