Dispose of used safety cartridges that 
are no longer required for the current 
application (see chapter “Disposal of 
equipment” on page

 

31).

 

!

The  safety  cartridges  are  supplied 
in  sterile  packaging.  Do  not  remove 
them  from  their  packaging  until 
immediately before you intend to use 
them.

 

!

Always  place  the  SOL  Nova  safely 
so  that  the  safety  cartridge  does 
not  come  into  contact  with  other 
surfaces.

 

!

Seek  medical  attention  if  injured  by 
a

 

contaminated needle.

 

!

Switch o

ff

 the power supply to the

 

SOL 

Nova.

 

!

Insert 

the 

safety 

cartridge 

in 

the

 

opening in the handle, as illustrated 

below.

Due to its locating points on the top and 
bottom, each safety cartridge can only 
fit in the oval opening in two ways.

 

!

Rotate  the  safety  cartridge  clockwise 
through approx. 45°.

As you rotate it, several locating points 
will engage audibly until the safety car-
tridge is firmly locked into position.

 

!

Check that the safety cartridge is firmly 
seated.

 

!

To remove the safety cartridge, turn it 
anti-clockwise and remove it from the 
handle.

5.3 

Connecting the power unit

 CAUTION

Risk of short circuits

If  cable  or  cable  connection  is  visibly 
damaged, there is a risk of damage to 
the electronics.

 

!

Perform  a  visual  inspection  of  the 
SOL Nova and the cable for signs of 
damage,  e.g.  a  defective  cable  con-
nection.

 

!

Never  bend  the  connecting  cable 
while in operation.

 

!

Observe  the  operating  instructions 
for the particular power unit.

 CAUTION

Cables constitute a trip hazard

People  can  stumble  over  inappropri-
ately routed cables and get injured.

 

!

Route  all  cables  so  that  no-one  can 
stumble  over  them  or  pull  them  ac-
cidentally.

ca. 45°

Cheyenne SOL Nova

Summary of Contents for Sol Nova

Page 1: ...sung Operating instructions Gebruiksaanwijzing Instructions Istruzioni per l uso Instrucciones de uso Manual de instruções Oδηγίες Opskrift Bruksanvisning Ohjeet Instrukcja obsługi Instrukce Navodila nt pl za uporabo Utasítás инструкции 说明 ...

Page 2: ...1 Cheyenne SOL Nova 1 1 2 3 4 5 ...

Page 3: ...gem Pega Motor Cabo de conexão Adaptador de cabo EL Safety cartridge Λαβή Μοτέρ Καλώδιο σύνδεσης Αντάπτορας καλωδίου DA Safety cartridge Gribestykke Drev Tilslutningskabel Kabeladapter SV Säkerhetspatron Grepp Motor Strömkabel Kabeladaptrar FI Safety cartridge turvapatruuna Kahvaosa Moottori Liitäntäjohto Johtosovitin PL Kartridż bezpieczeństwa Uchwyt Napęd Kabel zasilający Adaptery do kabla CS Be...

Page 4: ...cting the power unit 26 5 4 Fitting the protective hose 27 5 5 Checking equipment 28 6 Using the SOL Nova 28 6 1 Setting the needle protrusion 28 6 2 Setting the piercing frequency 29 6 3 Loading with ink 29 7 Cleaning and maintaining the SOL Nova 29 7 1 Material compatibilities 29 7 2 Disinfecting surfaces 30 7 3 Cleaning surfaces 30 7 4 Cleaning handle in an ultrasonic bath 30 7 5 Sterilising ha...

Page 5: ... notices Warning notices draw attention to the risk of personal injury and damage to materials and are structured as follows SIGNAL WORD Type of hazard Consequences Remedy Element Meaning Indicates a risk of injury Signal word Specifies the severity of the hazard see following table Type of hazard Gives the type and source of the hazard Conse quences Describes potential conse quences of non compli...

Page 6: ... switch it off and place it safely so that it cannot roll away and fall to the ground Note the technical data quoted in these operating instructions and comply with operating transport and storage conditions see Page 24 Hand over the SOL Nova to a specialist retailer for inspection if it exhibits any external signs of damage or if it does not function in the usual manner 2 3 Important hygiene and ...

Page 7: ...attooed Scars in the area that shall be tattooed Systemic infections and infectious dis eases e g Hepatitis type A B C D E or F HIV infection or acute local skin infections e g herpes rosacea During chemotherapy radiotherapy or high dosage corticosteroid therapy recommendation from four weeks be fore the start until four weeks after the end of the therapy Up to twelve months after a plastic sur ge...

Page 8: ...nowledge of the risks and side effects see chapters Contra indications on page 22 and Side effects on page 22 2 7 Intended use The SOL Nova is a professional tattooing machine for the application of tattoos to the human skin in dry clean and smoke free environments and under hygienic conditions It must be prepared used and maintained in the manner described in these operating instructions In parti...

Page 9: ... 5 to 12 6 V DC Power intake 3 W Start up current Max 2 5 A for max 200 ms Piercing frequency 25 to 150 Hz Stroke 3 5 mm Needle protrusion 0 to 4 0 mm Electrical connection 3 5 mm jack Drive Brushless DC motor Operating mode Continuous operation Diameter 33 mm Length 102 mm Weight Approx 150 g Sound emission pressure level Max 70 dB A Total vibration Max 2 5 m s 4 2 Operating conditions Ambient te...

Page 10: ...ecommended disinfectants in chapter Material compatibilities on page 29 Switch off the power supply to the SOL Nova Check if the SOL Nova has been contaminated badly by the return flow of ink or bodily fluids In such cases perform the steps described in chapter Cleaning and maintaining the SOL Nova see Page 29 Wipe down the connecting cables the drive and the handle with a soft cloth dipped in dis...

Page 11: ...the safety cartridge clockwise through approx 45 As you rotate it several locating points will engage audibly until the safety car tridge is firmly locked into position Check that the safety cartridge is firmly seated To remove the safety cartridge turn it anti clockwise and remove it from the handle 5 3 Connecting the power unit CAUTION Risk of short circuits If cable or cable connection is visib...

Page 12: ... of the connecting cable supplied to the connection on the drive unit of the SOL Nova Plug the jack of the connecting cable fully into the output socket on the power unit If you are using an third party power unit connect up its terminals correctly as described in the operating instructions of the third party power unit Connect the power unit to the mains power supply Cover the Power Unit or the t...

Page 13: ...pecialist retailer 6 Using the SOL Nova 6 1 Setting the needle protrusion CAUTION Risk of injury from excessive protrusion of needle A large needle protrusion permits a deep piercing depth If needle protrusion and piercing depth are excessive this can damage the subcutaneous layer Start with just a short length of needle protruding Switch on the power Adjust the needle protrusion while the SOL Nov...

Page 14: ... of the safetycartridgeinthedesiredcolourofink for 2 to 3 seconds with the unit running Avoid contact with the ink container The ink is picked up by the needles Alternatively the ink can be filled into the opening on the safety cartridge using a sterile pipette 7 Cleaning and maintaining the SOL Nova CAUTION Risk of short circuits If components of the SOL Nova are dis mantled and cleaned while sti...

Page 15: ... soft cloth dipped in a cleaning agent or a disinfectant Clean inaccessible surfaces e g the thread of the drive unit with a soft bristled nylon brush 7 4 Cleaning handle in an ultrasonic bath If the handle is severely soiled Use the cleaning agent Tickomed 1 or Tickopur R33 from manufacturer Dr H Stamm GmbH in accordance with the manufacturer s instructions Make sure that a pH value of 7 is not e...

Page 16: ...able regulations for used electronic equipment If necessary consult the specialist retailer or the responsible authorities to learn about applicable regulations Provide containers at the workplace for legislation compliant disposal of contaminated components Dispose of used or defective safety cartridges in a sharps container safety box in accordance with the regulations applicable in your country...

Page 17: ...al and workmanship for the product For the safety cartridges we guarantee sterility until the expiry date on the packaging providing that the packaging is sealed and undamaged and that the transport and storage conditions have been complied with For complaints relating to safety cartridges please advise us of the batch number printed on the label We do not provide warranty cover in respect of the ...

Page 18: ...ine directive Basic Machine directive requirements used or complied with 1 1 2 1 1 3 1 1 5 1 1 6 1 2 2 1 3 1 1 3 2 1 3 3 1 3 4 1 3 7 1 5 1 1 5 2 1 5 4 1 5 5 1 5 6 1 5 8 1 5 9 1 5 11 1 5 13 1 6 1 1 6 3 1 7 1 1 1 7 4 1 7 4 1 1 7 4 2 1 7 4 3 2 1 2 1 1 2 1 2 2 2 1 2 2 1 1 The special technical documentation according to Appendix VII Part B has been established The person authorized to compile the tech...

Page 19: ...0 3 3 ed 2 Electromagnetic compatibility EMC Part 3 2 Limits Limits for harmonic current emissions Electromagnetic compatibility EMC Part 3 3 Limits Limitation of voltage changes voltage fluctuations and flicker in public low voltage supply systems for equipment with rated current 16 A per phase and not subject to conditional connection CISPR 14 2 1997 AMD1 2001 AMD2 2008 CISPR 14 1 2005 AMD1 2008...

Reviews:

Related manuals for Sol Nova