25
EN
5
Getting the SOL Nova ready
for operation
ATTENTION
Damage from condensate
If the SOL Nova is exposed to significant
variation in temperature, e.g. during
transport, condensate can form inside
which can damage the electronics.
!
Ensure that the SOL Nova has
reached
ambient
temperature
before using it. If it was exposed to
high temperature fluctuations, wait
for at least 3 hours for every 10°C
of temperature di
ff
erence before
putting it into operation.
!
Only use the SOL Nova at ambient
temperatures of +10 to +35°C.
5.1
Disinfecting equipment
ATTENTION
Damage to drive caused by fluid
If disinfectant gets inside the drive, this
can corrode the electrical and mechan-
ical components.
!
Never immerse the drive in disinfectant.
ATTENTION
Product damage caused by non-
approved disinfectants
Disinfectants that are not compatible
with the materials of the product can
damage its surface.
!
Always use disinfectants that are
approved, i.e. that comply with
stipulations in your country.
!
Note the list of recommended
disinfectants in chapter “Material
compatibilities” on page
29.
!
Switch o
ff
the power supply to the SOL
Nova.
!
Check if the SOL Nova has been
contaminated badly by the return flow
of ink or bodily fluids. In such cases,
perform the steps described in chapter
“Cleaning and maintaining the SOL
Nova” (see Page
29).
!
Wipe down the connecting cables, the
drive and the handle with a soft cloth
dipped in disinfectant.
5.2
Replacing the safety cartridge
WARNING
Risk of injury from needles
Injuries caused by contaminated
needles from the safety cartridges can
lead to the transmission of disease.
!
Always use genuine Cheyenne
safety cartridges for the SOL Nova.
The installed safety membrane on
these safety cartridges prevents the
ingress of ink into the drive, and also
prevents pathogens from entering
the appliance.
!
Never hold the safety cartridge by
its tip.
!
Switch o
ff
the SOL Nova before
changing the safety cartridge.
!
Ensure that when removing a used
safety cartridge, no ink remnants can
run into the handle.
!
Never push the needles out of
the
safety cartridge.
!
Check during the removal of a used
safety cartridge that the needles are
completely retracted into the safety
cartridge.
Cheyenne SOL Nova