Filter Base System
1.
Place the o-ring in the beveled edge of the filter base.
2.
Place the perforated plate in the filter base, beveled edge facing up.
3.
Place the frit (filter) on top of the perforated plate.
4.
Fix the centering ring on top of the frit and make sure it is pushed down on all edges evenly.
5.
Wrap the lower outlet adapters with NPT end with PTFE tape. Thread the adapter into the filter
base until it becomes snug. DO NOT OVER TIGHTEN.
6.
Attach the beaded pipe coupling to the lower outlet drain and the CG-1959-A-01 0-14mm drain
valve. Torque the beaded pipe coupling to 35in-lbs using a torque wrench.
PLEASE NOTE: MAXIMUM TEMPERATURE LIMIT FOR FILTER BASE IS 150°C
Centering Ring
Frit (Filter)
*Note: The Bottom
Impeller Should NOT
Rub on the Filter
Perforated Plate
Filter Base