15
COMMUNICATION
CONNECTING COMMUNICATION CABLE
Turn off power and then connect the accessory
cable between the COM port of PC and spanner.
UPLOADING RECORD DATA
Make sure the connection between PC and
spanner is normal.
Press
together to reset the spanner.
Change the spanner operation mode to “Send”.
(Refer to “Peak Hold Mode Recorded Value
Review ” section)
Use PC to start the uploader program.
In uploader program, first select the correct
COM port No.
Next, select the file path to save the uploaded
data.
Finally, press “upload” button to transmit the
torque records to PC.
The uploaded data is then shown on the column
and saved in the *.csv file. Use Microsoft Excel
to view *.csv file.
Precaution:
1. Communication function is only supported
on some models. Check the model no. and
its specification before using
communication function.
2. Do not insert the plug of communication
cable into torque wrench that does not
support communication function.
CAUTIONS:
Refer to the uploader program user
guide for the detail operations.