– 14 –
6.4 Tare
NOTE:
Press
ZERO
to reset the measured value. This will allow the gauge
to ignore any force currently applied to the sensor shaft. The ignored, or
tared, force is still counted with regards to the gauge’s overload condition.
Overload will occur when the force applied to the sensor shaft exceeds
the gauge’s rated capacity by 20%. Stop measuring and remove the load
immediately to avoid permanent damage to the gauge. Any measurements
taken in the overload range are not accurate.
1. Tare is automatically performed when the gauge is first powered on.
Statistics Data
1. When in memory mode,
UNIT
will cycle through the available
statistics data.
2. Each press of
UNIT
will switch between the following items: positive
maximum value, negative maximum value, positive minimum value,
negative minimum value, average value, standard deviation.
3. Press
MEM
to exit to the recorded measurement values.
4. Press
PEAK
key to output the recorded memory via RS-232C.
(See section 12).
NOTE:
Pressing MEM exits statistics mode and returns to memory blocks.