AUTO Setup (cont’d)
Autoset
function menu
Function
Setting
Acquire Mode
Adjusted to Sampling
Display Format
Y-T
Display Type
Set to Dots for a video signal, set to Vectors for
an FFT spectrum; otherwise, unchanged
Vertical
Coupling
Adjusted to DC or AC according to the input
signal
Bandwidth
Limit
Off (full)
V/div
Adjusted
VOLTS/DIV
adjustability
Coarse
Signal inverted
Off
Horizontal
position
Centre
S/div
Adjusted
Trigger type
Edge
Trigger source
Auto detect the channel which has the input
signal
Trigger slope
Rising
Trigger mode
Auto
Trigger
coupling
DC
Trigger holdoff
Minimum
Trigger level
Set to 50%