20
IMEG 1000N
Earth protection
terminal
Red light when lit: multiply
the reading by 10 to get the
result (on M
W
scale)
Range switch to be returned
to V AC position after use
Black
Red
Green indicator when lit signals
that the magneto is being
correctly cranked: in other words
showing the presence of a
voltage that may reach 1000 V
between the terminals.
Fold-away handle
Measurement terminals
Guard terminal
Blue
Safety bar which serves as
a rest for the left hand whilst
the handle is turned
SWITCH POSITION
500 V/ V AC *
RANGE
600 V AC
ACCURACY
± 3% of the end of scale value
INPUT RESISTANCE
< 100 k
W
FREQUENCY
45...450 Hz
PERMITTED OVERLOAD
600 V AC and DC
TYPE OF PROTECTION
Over capacity components
* If the switch is on the 250 V or 1000 V position the displacement of the needle indicates
a live circuit but does not correspond to the result of the reading.
Mechanical zero reset screw for the galvanometer
Measurements of AC voltage (IMEG 1000N)
V AC
AC voltage to
be measured
Switch on
V AC
position