Press the
TEST
button and hold it down until the
measurement displayed is stable.
If a voltage greater than 25 V is detected, pressing the
TEST
button has no effect.
The measurement is displayed on the main display unit
and on the bargraph.
The secondary display unit indicates the test voltage
generated by the instrument.
M
G
V
M
The instrument displays the programmed test voltage.
To modify the test voltage (between 10 et 100 V),
press the
key.
When the first digit blinks, you can change it using the
key. Press
to go to the
next digit and
to change it. Press
one last time to validate.
V
DC
VAR
G
V
V
AC
VAR
G
V
Use the leads to connect the device
to be tested to the terminals of the
instrument. The device to be tested
must not be live.
Pressing the
key, during the
m e a s u re m e n t , c h a n g e s t h e
secondary display unit to display
the current or the elapsed time.
2.3. .
INSULATION MEASUREMENT
V
k
Ω
OFF
Ω
M
Ω
VAR
Set the switch to the
M
Ω
position.
The
symbol indicates that the instrument is generating a hazardous voltage (> 70 V).
The measurement results can be thrown off by the impedances of additional circuits connected in parallel or by transient
currents.
TEST
> 1s