3.1. MEASUREMENT OF A VERY LOW VALUE
±I
Reverse the direction of the current by
pressing the ±I key and the instrument
displays the average:
R(+I) + R(-I)
2
T h i s s e r v e s t o e l i m i n a t e a n y
thermoelectric effect.
START/STOP
START/STOP
To display the
values R(+I) and
R(-I), press the
DISPLAY key.
R
AVG
=
...or disconnect one of
the 2 clips.
A
m
Ω
2nd
+I
HOLD
2
C
2
P
1
P
1
C
In both cases, the
last measurement
made is displayed,
along with the
HOLD
symbol.
If the measurement is stopped by the
disconnection of a clip, simply connecting
it to another object starts another
measurement, with no need to press the
START/STOP key.
A
AVG
HOLD
I
_
DISPLAY
mV
m
Ω
To display the voltage on the terminals of
the resistance instead of the measurement
current, press the DISPLAY key.
A
µ
Ω
AVG
HOLD
I
_
DISPLAY
+I
R(+I)
+V
R(+I)
-I
R(-I)
-V
R(-I)
The energy accumulated in an
inductive
component when a measurement is made on it must be released.
You must never in any circumstances touch or disconnect the measurement leads until you
have stopped making the measurement and waited at least ten seconds
for all of the energy in the
item tested to be dissipated. Failure to observe this precaution may result in the production of an arc,
potentially hazardous for both the instrument and the operator.