30
2. BATTERY CHARGE
Start by fully charging the battery before the first use.
In the 10 A range, the battery life is approximately 1h20. It is therefore best to charge the battery before
starting series of measurements. Charging must be done between 0 and 40°C.
The battery life of the instrument depends on the range. To display it (before making the
measurement):
DISPLAY
> 2s
SET-UP
400m
Ω
1A
4000m
Ω
100mA
40
Ω
10mA
400
Ω
10mA
40m
Ω
1A
4000µ
Ω
10A
OFF
Charge
OFF
Charge
OFF
Charge
Current battery
capacity in %.
Charging time: 3 h 30
Average
b a t t e r y l i f e
according to
range
Following long-term storage, the battery may be completely discharged. In this case, the first charge
may last several hours. The capacity of the battery and therefore the battery life of the instrument will
be temporarily reduced. The battery will recover its initial capacity after 5 recharging cycles.
Remaining battery life