The procedure to follow is as follows :
1. Connect the remote control pedal in accordance with fig 41.
2. Record a first measurement at the desired memory location (memory number and machine number) with the keys on the
keyboard (see § 5.2).
3. Carry out the next test by pressing the START/STOP pedal.
4. Save the results by pressing the SAVE pedal.
The following window (fig. 42) is displayed for a moment (if for example an HT test
has been carried out) :
Then the main menu is displayed again. Continue the measurements.
■
■
■
■
■
Technical specifications of the pedal :
Cable length : 10 m
Commands : START/STOP, SAVE
Case
: metallic
Weight
: 2 kg
Dimensions (W x H x D) : 300 x 175 x 55 mm
5.10. USE OF WARNING LAMPS (STANDARD VDE 104)
The warning lamps are used to warn the user and other persons who may be present on the sites of the tests of the presence of
dangerous voltages during dielectric tests.
■
■
■
■
■
Meaning of the colours :
- The red light (TEST)
lit
means the presence of a dangerous voltage at the terminals of the dielectric test. Please take care
during handling of the test probes (guns) ! DANGER PRESENT !
- The green light (READY)
lit
means that the instrument is powered up and ready for a measurement; there is not a dangerous
voltage at the terminals of the dielectric test. POTENTIAL DANGER !
Fig.43. Connection of the warning lamps to the C.A 6121 MACHINE TESTER
■
■
■
■
■
Technical specifications of the warning lamps:
Cable length : 2 m
Bulbs
: 12-15V / 4W
Case
: plastic
Weight
: 0.3 kg
Dimensions (W x H x D) : 200 x 110 x 95 mm
NOTE :
If no lamp lights up when the rotary switch is in the HV position, stop the measurements immediately and check the
connection of the warning lamps as well as the bulbs.
Fig.42. IMemory and machine
number information
HV
1000V 500mA
mA
20.3
MACH
No
MEM
MACH:127
No:028 FR
U:1.08 kV