How to perform the measurement ?

1. Set the rotary switch to 

t (discharge time) position; the screen in

fig.21 is displayed.

2. Select 4-pole system by pressing the SYST. key (4-point 5s is

displayed)

3. Connect the test cables to the instrument and to the tested unit in

accordance with figure 26; le 

 symbol is displayed above unit

S, informing the user that the mains voltage is connected to the
MAINS TEST input.

4. Press the START/STOP key to prepare the instrument for the mains

voltage cut-out; Ready is displayed after 1s approx.

5. Pull out the double connection element and wait for the result to be

displayed.

6. Save the displayed result for documentation purposes (refer to the

instructions in chapter 5.2 to save the displayed result).

  NOTES :

The red TEST ON lamp is on during the measurement (until pulling out the double connection element and until the TEST voltage
has dropped to 60 V).

The actual voltage is displayed during the measurement, and the disconnection voltage (see fig.20) is displayed when the
measurement is finished.

t < 0.1 s is displayed under the main result and no voltage is displayed on the right of the display, if the discharge time is less

than 0.1 s.

5. OPERATION

5.1.

WARNINGS

Different warnings in addition to information can be displayed during operation of the C.A 6121 MACHINE TESTER. See below a
list of warnings and information for each function.

HV function :

Trip out

: The HV generator tripped because the test current was higher than the set threshold value.

U function :

U > 

U

max

: The displayed result (

U) is higher than the set threshold value.

U > 12 V

: The AC voltage connected to the input P1-P2 is higher than 12 V.

R

10A

 function :

R > R

max

: The displayed result (R) is higher than the set threshold value.

U > 12 V

: The AC voltage connected to the input P1-P2 is higher than 12 V.

M

 function :

R < R

min

: The displayed result (R

ins

) is lower than the set threshold value.

U > 30 V

: The AC or DC voltage connected to the MW terminals is higher than 30 V.

t function :

Voltage?

: There is no voltage conected to MAINS TEST terminals.

t < 0.1 s

: The voltage at the terminals of the MAINS TEST inputs dropped below 60 V in less than 0.1s.

t > tmax

: The displayed result (

t) is higher than the set threshold value (chosen at the same time as the connection

system; 2 pole or 4 pole).

All functions :

HOT

: The instrument is overheated (in functions 

U, R

10A

 or HV) 

 wait.

Measurements in functions M

 or 

t can still be carried out.

NO RESULT

: There is no result on the display or the displayed result has already been saved.

NOTHING TO PRINT : There was no result saved/displayed before entering the PRINT menu.

Fig.26. Connection of the test leads

C

U :
U

:

t

:

R :

in
lim

lim

in

t:

D

0...600 V AC/DC
60 V DC
5 s
96 M

W

0...10 s

C2

P2

C1

P1

600V 250V CAT III

1 - 1.25 - 1.5 kV

START / STOP

TEST ON

MEM/MR

PRINT/RS

SET-UP

C.A 6121

HV

M

W

R

10A

D

t

D

U

ARNOUX

CHAUVIN

INSULATION (M )

W

500 - 1000 V

DISCHARGE TIME ( t)

D

CONTINUITY 12V/>10A

( U,R

)

D

10A

DIELECTRIC (HV)

L

N

PE

Summary of Contents for C.A 6121

Page 1: ...150M U 513V 9min 59s M 113 5 Timer Rmin Un n n n n n CONTROLEUR MACHINE n n n n n MACHINE TESTER n n n n n MASCHINEN TESTER n n n n n MACHINE TESTER n n n n n COMPROBADOR DE MAQUINARIA ELECTRICA User...

Page 2: ...es Use connection accessories whose overvoltage category and service voltaage are greater than or equal to those of the circuits on which the measurements are made Only use accessories that conform to...

Page 3: ...al specifications 32 4 MEASUREMENTS 32 4 1 Dielectric test 32 4 2 Voltage drop scaled to 10 A position U 34 4 3 Low resistance position R10A 35 4 4 Insulation resistance 36 4 5 Discharge time 2 pole s...

Page 4: ...of 1500 V AC Voltage drop scaled to 10 A AC with a test voltage of 12 V 10 A AC Low resistance with a test voltage of 12 V 10 A AC Insulation resistance with a test voltage of 500 V DC Insulation res...

Page 5: ...TECHNICAL SPECIFICATIONS 3 1 DIELECTRIC TEST POSITION HV Nominal test voltage adjustable 1000 1250 or 1500 V 50 Hz Test voltage 1000 V 1250 V or 1500 V 50 Hz at Umains 230 V and a power Pload 500 VA...

Page 6: ...nt 0 0 5 10 A AC Timer adjustable from 1 to 15 s resolution 1s or without timer Connection system 4 wire 3 3 LOW RESISTANCE POSITION R10A Resistance readout Range R Resolution m Accuracy 0 999 m 1 2 o...

Page 7: ...2 counts 20 0 199 9 100 200 500 1000 10 rdg automatic ranges Range 0 200 M stable result even on a capacitive load Nominal voltage 1000 V DC 10 0 Short circuit current 1 4 mA max Measurement current...

Page 8: ...Dimensions W x H x L 400 x 260 x 250 mm Weight 11 kg Protection index IP 40 open IP54 closed Working temperature range 0 50 C Reference temperature range 5 35 C Storage temperature range 10 60 C Max...

Page 9: ...bove the mA unit Press EXIT key to exit this menu Tripping out current fig 6 Press I key to open the tripping out current selection menu Press the or keys to choose the tripping current Press EXIT key...

Page 10: ...es equipped with longer cables 6 m When the test current is greater than the pre set limit the HT generator trips automatically when this value is exceeded and the pre set limit value is displayed as...

Page 11: ...er value Timer key to set timer value or to switch it on off Resistance main result Rmax key to set threshold resistance Fig 12 Basic screen in position R10A 2 Select the test parameters as follows Re...

Page 12: ...Basic screen in M position 2 Select the test parameters as follows Insulation resistance threshold Press the Rmin key to access the menu for insulation resistance threshold selection Press the ou keys...

Page 13: ...C A 6121 MACHINE TESTER OBJECT UNDER TEST MAINS TEST Fig 19 Test circuits discharge time 60 V t expected MAINS TEST input voltage disconnection voltage ideal power disconnection moment power disconnec...

Page 14: ...voltage see fig 20 is displayed when the measurement is finished t 0 1 s is displayed under the main result and no voltage is displayed on the right of the display if the discharge time is less than...

Page 15: ...on for each function HV function Trip out The HV generator tripped because the test current was higher than the set threshold value U function U Umax The displayed result U is higher than the set thre...

Page 16: ...OTES A displayed result can only be saved once in order to avoid the risk of storing twice Memory save locations already occupied can be erased by saving a new result at this location The occupied loc...

Page 17: ...NE TESTER 9 pin male connector to serial printer 2 2 3 3 5 5 front view front view Fig 29 RS232 communication cable printer C2 P2 C1 P 1 TEST MAINS TEST REMOTE CAT III 600V WARNING LAMP 12V 1 1 25 1 5...

Page 18: ...rs 001 007 in our example will be printed when you press the START key Press the CHG key then the keys to modify the machine numbers Print all All the results will be printed when you press the START...

Page 19: ...re of the PC to communicate with the PC Pressing the PRINT RS exits the RS 232 mode and opens the main menu Communication with the PC is then not possible NOTES Check the RS 232 format in chapter 3 6...

Page 20: ...1 9 V 2 5 mm2 1 4 V 4 0 mm2 1 0 V 6 0 mm2 t Timer 1 s 15 s 10 s R 10A AC Rmax Max permitted 10 m 1000 m in steps of 10 m m resistance or m no limit t Timer 1s 15 s 10 s UN Test voltage 500 1000 V DC...

Page 21: ...time set up menu is not running 5 7 3 Warning buzzer One of the functions of the warning buzzer is to provide audible information relating to the evaluation of the measurement result set threshold val...

Page 22: ...III 600V TEST MAINS TEST REMOTE WARNING LAMP 12V 1 1 25 1 5 kV START STOP TEST ON 230V 50Hz 500VA RS232C MEM MR PRINT RS SET UP C A 6121 MACHINE TESTER ON OFF HV MW R10A Dt DU ARNOUX CHAUVIN INSULATI...

Page 23: ...USE OF WARNING LAMPS STANDARD VDE 104 The warning lamps are used to warn the user and other persons who may be present on the sites of the tests of the presence of dangerous voltages during dielectric...

Reviews: