2

Neo Stimulation Channel 1 & 2 + EMG Module  Manual

Install module from the left side (when facing the screen)  

of the unit. 

The Neo Stimulation Channel 1 & 2 + EMG Module is 

designed only for use with Vectra Neo or Intelect Neo 

Therapy Systems.

Items included in the modules:

Electrotherapy Module Channels 1/2 – PN 70000

•  Stimulation module

•  Lead wires

•  DURA-STICK® 2 in (5 cm) Round Disposable Electrodes 

(1 pack of 4)

•  Faceplates (to cover module after inserted)

Electrotherapy Module Channels 1/2 + sEMG – PN 

70004

•  Stimulation module (2 channel Stim with sEMG)

•  sEMG Leadwires

•  DURA-STICK® 2 in (5 cm) (2 packs of 4) Round 

Disposable Electrodes

•  Faceplates (to cover module after inserted)

Tools required (not included): 

•  #2 Phillips screwdriver 

•  Standard slotted screwdriver

The system is programmed to automatically recognize 

the new Module, therefore, no software installation is 

required.

Complete the following steps to install the module:

1.  Power off device. Remove main power cord from the 

rear of the device.

2.  Remove the blank faceplate over the second slot from 

the front on the left and right sides of the unit. Insert 

a standard slotted screwdriver in the top slot, pressing 

down with slight pressure. Pull the faceplate away.  

3.  The second slot is illustrated below, but the technique 

is the same for all modules.

 

4.  The module is inserted on the left side in the second 

slot as shown. Carefully insert the module into the 

slot, with 32 pins (2 x 16) in first. 

5.  Set the module in place with gentle pressure until you 

feel an end point of travel.

MODULE INSTALLATION

Summary of Contents for Neo Stimulation Channel 1 & 2

Page 1: ...el 1 2 EMG Module 13 8905 Rev F For complete User Operating Instructions including Cautions Warnings Dangers Indications and Contraindications refer to the User s Manuals Domestic Market US only refer toVectra Neo ClinicalTherapy System User Manual Item 13 7646 CDVersion item 13 7647 International Market refer to Intelect Neo ClinicalTherapy System User Manual Item 13 7651 CDVersion item 13 7652 T...

Page 2: ... required not included 2 Phillips screwdriver Standard slotted screwdriver The system is programmed to automatically recognize the new Module therefore no software installation is required Complete the following steps to install the module 1 Power off device Remove main power cord from the rear of the device 2 Remove the blank faceplate over the second slot from the front on the left and right sid...

Page 3: ...fer to Intelect Neo Clinical Therapy System User Manual Item 13 7651 CD Version item 13 7652 To obtain a copy contact your local representative or DJO Customer Care 6 Using a 2 Phillips screwdriver secure the module with the provided 8 32 x 5 16 screw at the bottom tab as shown 7 Insert the faceplate with connection openings supplied with module on the Left side at the bottom and snap into place a...

Page 4: ... ECG monitors and ECG alarms may not operate properly when electrical stimulation is in use TENS is a symptomatic treatment and as such suppresses the sensation of pain which would otherwise serve as a protective mechanism Contaminated electrodes and leadwires can lead to infection Use of electrode with degraded hydro gel can result in infection to the skin burn Use of electrode on multiple patien...

Reviews: