Intelect® Transport 2-Channel Electrotherapy
45
OPERATION
Restoring Factory Protocols
If necessary, you can choose to restore the unit’s original (default) waveform parameters when it was shipped to you.
NOTE:
This procedure will erase all user-defined protocols.
To restore the unit’s original waveform parameters, do the following:
1.
Press
the
Clinical
Resources
button.
The
Clinical
Library
window
displays.
2.
Press the Up Arrow or Down Arrow buttons to highlight the Restore Factory
Protocols
option.
3.
Press the Enter button to accept the highlighted selection.
The unit displays the message "Restoring Protocols Please wait."
The user-defined protocols are erased and the waveforms are restored to the original
parameters.
Then
the
Restore
Factory
Protocols
Confirmation
window
displays.
4.
Press any button on the Operator Interface.
You
are
returned
to
the
Clinical
Library
window.
SYSTEM UTILITIES continued