28
Vectra Genisys® Laser Module
SETUP
REMOVING THE MODULE FROM THE UNIT
To remove a module from the main unit, do the following:
1.
Disconnect the unit from the electrical outlet.
2.
Remove the rear access panel and disconnect the power cord from the main unit.
3.
Remove the extended front access panel and disconnect all cables and wires.
4.
Route the Power Cord through the Power cord routing area on the Module to the unit.
5.
Replace the Rear Panel.
6.
Using a #1 screwdriver, remove the four screws that are securing the module. Retain the screws for future use.
7.
Remove the module from the unit.
8.
Using only your fingers, gently remove the female end of the ribbon cable from the connector pins of the module.
9.
Leave the ribbon cable connected to the main unit.