Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
31
6 REMOVAL & REPLACEMENT
A. Part Numbers
Gray Applicator Cable
..................................... 27351
Blue Applicator Cable
..................................... 27106
B. Tools and Equipment Required
#1 Phillips Screwdriver
C. PC Board Removal and Replacement
1. Using a counterclockwise motion, turn the
end of the Applicator to remove the
Applicator
Cable.
Refer to Figure 6.29.
2. Once the end of the cable is free from the
Applicator housing, grasp the wires of the
cable and gently pull the applicator
connector from the holder on the PC Board.
3. To install position the two black tabs of
the replacement Cable away from the inside
of the Applicator. Using a small screwdriver
push the Connector onto the pins of the
Connector Holder. Gently push the sides of
the Connector to verify that it is well seated.
4. Once the Cable Connector is well seated,
hold the Cable and twist the wires 2 1/2
turns
counterclockwise.
Refer to
Figure
6.30.
NOTE:
Twisting the Cable wires prior to screwing
the Cable onto the Applicator housing
will allow the wires to untwist and
straighten out when the Cable is screwed
into
place.
5. Verify the operation of the Applicator.
•
6.11 APPLICATOR CABLE
FIGURE 6.29
FIGURE 6.30