5 TROUBLESHOOTING
5.7 ULTRASOUND BOARD TEST
NOTE:
This test is designed to eliminate or verify a
problem with the Ultrasound Board.
A.
Test
1. Separate top from bottom.
Refer to 6.1,
part
C
.
2. Disconnect the Power Supply Harness from
the Control Board. Lay the unit open.
Refer to Figure 5.2.
3. Remove the Ultrasound Board.
Refer to 6.4,
steps 2 and 3
.
NOTE:
The Ultrasound/Control Board Header may
stay attached to either the Ultrasound Board
or the Control Board. It may be necessary to
remove the Header for installation.
4. Place the top of the unit back on the bottom
of the unit. While holding the top and
bottom together, with the Ultrasound Board
missing, plug the unit back into the Mains
Power.
5.
Press the power button to verify that the unit
turns
on.
NOTE:
The Power On/Off button will flash when
plugged into Mains Power if the Ultrasound
Board is not functioning properly.
B. Test Results
1. Unit will power up, unit passed test.
a) Replace the Ultrasound Board.
2. Screen does not display, unit failed test.
a) Contrast Control needs adjusting.
b) Possible bad Display.
c) Possible bad Control Board.
d) Possible bad Power Supply.
FIGURE 5.2