8
Installation
Product Manual - CryoDoser FleX
™
LN
2
Dosing System
2
Select a nozzle.
3
Insert the nozzle into the nozzle tool, threads out (see
Image 9).
4
Thread the nozzle into the dosing head area in a
clockwise direction (Image 10 & 11). Do not over
torque
5
Re-apply the dosing head heater.
Caution! Never use any sharp or pointed
object such as an ice-pick,
screwdriver, torch or similar
device on the dosing head
(Image 10). The ribs of the
internal bellows are a thin walled
metal and the hole on the outer
ring of the dosing head is a
positive pressure port to help
keep moisture out and ice from
forming. High heat and puncture
holes will destroy the vacuum
insulation and VOID WARRANTY.
Image 9
Image 10
Image 11