Gross Particle Filters:

There is a gross particle (foam) filter located at the bottom of the device at the filter door.  This filter does not require any patient maintenance 
or cleaning, but should be replaced every 2 years as part of preventative maintenance or more frequently as needed.  

An OPTIONAL gross partcile filter can also be used in the handle of the device.  This is not required for proper operation, but can be utilized 
for harsh environments or provider requirements.  If this is in use, the patient should inspect and clean it once weekly.  

HEPA Filter

The HEPA filter is located inside of the Companion 5 on the front case half of the unit. It is designed to last the life of the unit and does not 
have a requirement for replacement.

Preventative Maintenance

Preventative Maintenance (PM) should be performed every 2 years. It can be performed more frequently as required by the provider based on 
their requirements or dirty operating environments. Preventative Maintenance consists of:
 

• Replace the Air Intake Filter

 

• Replace the gross partcile filter at the bottom of the device.

Maintenance should not be performed by patients. Note: Step-by-step instructions for these maintenance procedures can be found in the 
Provider Technical Manual.
  

Oxygen Concentration 

The Companion 5 will produce oxygen at 90% (+5.5/-3%) at all flow rates. 

Oxygen concentration should be tested upon delivery to a patient and at periodic intervals determing by the equipment provider. 

Hour Meter

The LCD hour meter is clearly visible to the right of the flow meter on the front of the Companion 5. It displays the cumulative hours to the 
nearest tenth and cannot be re-set. 

The hour meter will display a diagnostic alarm code if the Companion 5 is in an alarm condition. 

Alarms

Condition

Visual Indicator

Audible Indicator

Alarm Code

None

Green Solid Light

None

None

System Malfunction

Red Solid Light

Continuous Alarm

AL-P01

Power Loss

Red Solid Light

Continuous Alarm

AL-P02

Low Flow

Red Flashing Light

Beeping

AL-P20

High Flow

Red Flashing Light

Beeping

AL-P40

Low O2 Concentration (<85%)**

Yellow Flashing Light

Beeping

AL-P08

Low O2 Concentration (<70%)**

Yellow Solid Light

Continuous

AL-P04

**Only for OCSI units**.

autoFLOW

 Technology

Companion 5 features a patent-pending variable valve timing system that adjusts the valve cycle timing based on flow rate.
 

• Reduces power below 300W @ 3.0LPM below.

 

• Reduced operating temperatures, system pressures, sound levels, and component strain at lower flow rates. 

 

• Reduces compressor lead by 25%