Chart Environmental Chambers 

 

System Installation 

suction into the blower in the back wall of the chamber and discharge from the top of the 
chamber. 
 

 

❏ 

Liquid Nitrogen Plumbing 

 
Along with the electrical power, bring in the LN

2

 service from the top or the back of the 

chamber for easy access around the chamber.  The connection point is at the solenoid 
valve on the back wall of the chamber.  The nitrogen then flows through the valve and 
then enters the chamber to supply liquid nitrogen to the nozzle inside the plenum.  The 
location of the liquid nitrogen inlet is dimensioned on the O&D drawing at the end of this 
manual. 
 

 

Liquid nitrogen 
solenoid valve 

Gaseous nitrogen 
exhaust connection. 

 

A line relief valve should be installed between the LN

2

 connection on the chamber (the 

solenoid valve) and the first manual shut-off valve upstream from this connection in the 
nitrogen pipe system.  This relief valve should be rated at a pressure less than 175 psig, 
or the lowest pressure rated component in the system.  This relief valve is not included 
with the system and should be taken care of at installation. 
 

❏ 

Exhaust Gas Plumbing 

 
The exhaust gas of the chamber should be plumbed to the outside of the building.  No 
fans or restrictions should be placed in this line.  Any fans or restrictions will decrease 
system performance.   
 
It is important to vent the system to the outside of the building for safety reasons.  If the 
system is not vented to the outside of the building, there is a risk of an oxygen-depleted 
atmosphere in the area surrounding the chamber. 
 
The vent piping should also be well insulated to prevent moisture present in the air from 
condensing/freezing on the outside surface of the exhaust pipe.  The exit of the exhaust 
piping should be configured to prevent any water, rain, etc from flowing back into the 
chamber’s exhaust system.  Any water present in the system may form ice and break the 
cooling heat exchanger. 
 

 

 

8

Summary of Contents for 11744195

Page 1: ...Chart Test Chambers CHART INDUSTRIES INC ISIS CHAMBER USER S MANUAL Rev A...

Page 2: ...N 55306 USA Design Modification DO NOT use this product in any manner not consistent with the instructions outlined in this manual NEVER alter the design or perform service that is not consistent with...

Page 3: ...ilure prone components and assemblies in the shortest possible time The unique characteristics of Chart Inc Environmental Chamber systems are of key importance in achieving this effectiveness The staf...

Page 4: ...this symbol needs extra attention NOTE Text in this format is extra information helpful to the situation CAUTION Text in this format is information to help avoid personal injury and or property damage...

Page 5: ...RUMENTS 5 TEMPERATURE SAFETY CONTROLLER 6 SYSTEM INSTALLATION 7 LOCATION 7 ELECTRICAL POWER HOOK UP 7 LIQUID NITROGEN PLUMBING 8 EXHAUST GAS PLUMBING 8 OPERATION 10 OPERATOR CONTROL SWITCHES 10 WATLOW...

Page 6: ...ntents List of Tables TABLE 1 CHAMBER PERFORMANCE SPECIFICATIONS 4 TABLE 2 SYSTEM DETAILS 4 TABLE 3 DIMENSIONS AND UTILITIES 4 TABLE 4 LIQUID NITROGEN COOLING VALVE PARTS 16 TABLE 5 REPLACEMENT THERMO...

Page 7: ...peration Servicing of this equipment and its associated utility service must be done only by properly trained personnel Liquid nitrogen and high voltages can all cause severe injury or death if not pr...

Page 8: ...fety system A separate thermocouple is provided for the limit control and located inside the air plenum If this device has reached an alarm condition the system must be reset by pressing the Chamber R...

Page 9: ...ortional solid state control of resistive heating elements in air conditioning plenum Programmable Control and Data Logging Controller Watlow Series 96 temperature controller Communications A serial c...

Page 10: ...ing environment around the chamber to ensure safe operating conditions The liquid nitrogen control is done by a solenoid valve The solenoid valve is chosen for its compact size and high reliability Th...

Page 11: ...Controller A temperature Hi Lo limit control protects the chamber against thermal runaway due to a primary controller fault or control device failure The input to the Hi Lo limit control is a type T...

Page 12: ...k of the control console for the entrance of the power wires The power specs are given in the Specs section of this manual It is extremely important to run a properly sized ground wire It is often ben...

Page 13: ...he nitrogen pipe system This relief valve should be rated at a pressure less than 175 psig or the lowest pressure rated component in the system This relief valve is not included with the system and sh...

Page 14: ...the room containing the chamber as well as the ventilation system number of air changes must be done to determine this If it is determined that the room is ventilated well enough to not plumb the exh...

Page 15: ...e requirements are listed in the Specifications section 2 With all function switches in OFF position connect power to the chamber 3 Be sure chamber enclosure door is closed securely 4 Turn the Control...

Page 16: ...chamber from thermal run away and extreme temperature conditions 5 SERIAL COMM communications connection This connector is provided on the back of the control box for future serial communications wit...

Page 17: ...ure Set Points The temperature set point for the chamber is displayed on the bottom display of the Watlow controller The set point is simply entered by using the up and down arrow buttons on the front...

Page 18: ...l is a shielded type T thermocouple The thermocouple is mounted in the air flow plenum and is not directly visible The temperature safety controller has two independent set points One set point is a h...

Page 19: ...t the LN2 supply is on and that there is sufficient LN2 to perform your test The pressure of the LN2 should also be verified and maintained at less than 175 psig Maintenance Tasks to be Performed on a...

Page 20: ...m and tighten if necessary Maintenance Tasks to be Performed Every Three Months 1 Measure and record the resistance of the heater Changes in these measurements over time can help predict any future he...

Page 21: ...rugged and shielded sensor designed to withstand the entire temperature range of the chamber The shielding helps ensure reliable input signals to the temperature controller Chart sells this wire by t...

Page 22: ...lation thereof will be the responsibility of the customer Accessories and components not manufactured by Chart Industries Inc shall be warranted to the extent of the original manufacturer s warranty C...