Maintenance and Calibration
The Jewel
®
Workstation Continuous Mini Monitor is
solid state and designed to be maintenance free. It
is calibrated to NIST traceable standards. There are
no user adjustments necessary. Because of the wave
distortion impedance sensing nature of the test circuit,
special equipment is required for calibration. We
recommend that calibration be performed annually
using the Charleswater
Continuous Monitor
Calibration Unit. The Calibration Unit is a most important
product which allows the customer to perform NIST
traceable calibration on continuous monitors. The
is designed to be used on the shop floor at the
workstation, virtually eliminating downtime, verifying that
the continuous monitor is operating within tolerances.
provides a simple and quick means to assure
compliance verificaton Per ANSI/ESD S20.20 Table 3
per ESD TR53.
Limited Warranty
Charleswater expressly warrants that for a period of one (1) year
from the date of purchase, Charleswater Continuous Monitors
will be free of defects in material (parts) and workmanship
(labor). Within the warranty period, the product will be tested,
repaired or replaced at Charleswater’s option, free of charge. Call
Customer Service at 00 44 (0) 1892-665313 for a Return Material
Authorisation (RMA) and for proper shipping instructions and
address. You should include a copy of your original packing slip,
invoice, or other proof of purchase date. Any unit under warranty
should be shipped prepaid to the Charleswater factory. Warranty
replacements will take approximately two weeks.
Warranty Exclusions
THE FOREGOING EXPRESS WARRANTY IS MADE
IN LIEU OF ALL OTHER PRODUCT WARRANTIES,
EXPRESSED AND IMPLIED, INCLUDING
MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE WHICH ARE SPECIFICALLY DISCLAIMED.
The express warranty will not apply to defects or damage
due to accidents, neglect, misuse, alterations, operator
error, or failure to properly maintain, clean or repair
products.
Limit of Liability
In no event will Charleswater or any seller be responsible or
liable for any inury, loss or damage, direct or consequential,
arising out of the use of or the inability to use the product.
Before using, users shall determine the suitability of the
product for their intended use, and users assume all risk
and liability whatsoever in connection therewith.
2. Remove the charges from your body by making
contact with a properly grounded ESD worksurface.
Snap the cord to the wristband, and fit it snugly onto
the wrist. This should silence the audio and cause
the LEDs to switch from red to green. If this does not
happen, check the coil cord for continuity or damage.
Examine the wrist and secure that it has a secure
fit. If you have dry skin, apply an approved dissipative
hand lotion such as Menda
35666
Reztore™ ESD
Hand Lotion. When leaving the area, a person can
take the coil cord along or leave it attached to the
monitor in park position.
NOTE:
Removing the banana plug from the monitor’s
operator jack will disable the audio alarm.
It is recommended that you use the Park Snap rather
than continuously removing the Banana Plug from its
jack. Doing so will increase the life span of the operator
jack. If your process requires the constant insertion
and removal of the banana plug, please contact your
Charleswater Representative for alternate solutions.
Specifications:
Test range of monitored circuit:
500K - 10M ohms†
Worksurface Limit*:
Set to 500 Megohms ± 20%
Operating Voltage:
24VAC, 50-60 Hz
Wrist strap open circuit voltage:
1.2 Volts peak to peak @ 1-2 MicroAmps
Mat test open circuit voltage:
5 - 7.5 Volts
Response time to alarm:
<50 mS
Operating Temperature:
0ºC - 38ºC
Humidity Range:
0-95% RH, non- condensing
Dimensions:
70.4mm L x 52.6mm W x 18mm H
† This cannot be verified with standard DC test
equipment. The continuous monitor is an impedance
sensing device and the limits are determined by the
magnitude and angle of the impedance.
*Contact factory for special limits.
NOTE:
Worksurface must have a conductive layer such
as Dual Layer Rubber or Dissipative 3-Layer Vinyl or
Micastat
®
Dissipative Laminate with conductive buried
layers. Charleswater continuous monitors are not
recommended for use with homogeneous matting.