background image

PPE-5075.E

 Page 3 of 6

© 2012 CHARLESWATER LTD.

Unit 17. Millbrook Business Park, Sybron Way • Crowborough, East Sussex TN6 3JZ United Kingdom

Phone: 00 44 (0) 1892-665313, Fax: 00 44 (0) 1892-668838 • Internet: 

Charleswater.co.uk

Cleaning

Clean the conductive pads of electrodes with a minimum 70% 

isopropanol-water solution.”  Make sure they are dry prior to use. 

See specific product test standard for laboratory test specimen 

cleaning instructions. Generally, the test specimens and electrodes 

should be cleaned twice with a minimum 70% isopropanol-water 

solution using a clean, low-linting cloth each time.” (Note: then 

conditioned for 72 hours, minimum)

For installed product periodic audits, do not clean surfaces.  

However, if any measurements lie outside the acceptable range, 

clean the surface and re-test.  (Note: for working surfaces, use 

Reztore™ Antistatic Surface and Mat Cleaner (item 71021) or 

other ESD cleaner not containing silicone. Be sure the surfaces dry 

before testing.

Periodic maintenance - The area surrounding the cable jacks at the 

top end of the meter should be wiped with a clean cloth moistened 

with alcohol to remove skin oils that will accumulate and affect 

the accuracy at high resistances.  The frequency of cleaning will 

depend on usage;  once a month would be a good starting point.  

Other items that should also be cleaned in this fashion are the 

cable jackets and the resistivity attachment, if included.

Power Requirements

The Meter is powered by two replaceable alkaline AA batteries.

Test Procedure

General Guidelines:

•  Use both 2.27 kg Electrodes for Rg

•  Use one 2.27 kg Electrode with lead to grounded EPA 

bonding point for Rp (note: EPA bonding points are usually 

snaps installed on the material or workstation)

•  Use optional Resistivity Attachment (removing leads & 

Electrodes) for Resistivity measurements

•  Ensure that item being measured is electrically isolated (i.e. 

placed on an insulative surface) or Meter may measure lower 

resistance path.  Lay item on an insulating support (>10E12 

ohms) avoiding contact with the table.

•  Ensure that test leads are separated or Meter may measure 

lower resistance path

•  When using 2.27 kg Electrodes:

  -  Place no closer than 5 cm from edge of surface being 

measured

  -  Place no closer than 7 cm to any EPA bonding point

  -  Place 2.3 kg electrodes about 25cm apart for Rp-p

  -  Preferred placements include: most commonly used 

surface portion, most worn, center, furthest from EPA 

bonding point.

  -  For Rg, connect the sensing lead with shielded plug to 

grounded EPA bonding point

•  If surface has sections (like floor tiles or garment panels), for 

Rp place a 2.27 kilogram electrode on different sections

•  Clean surface for Laboratory test, but do not initially clean 

surface for installed products (if fails, clean and retest)

Laboratory Test Procedure Guideline

For laboratory test of ESD Working surfaces, Floor Materials, 

Footwear, Garments, or Seating, best advice is to follow test 

methods of Annex A of EN 63140-5-1.

•  The test specimens and electrodes should be cleaned twice 

with a minimum 70% isopropanol-water solution using a 

clean, low-linting cloth each time)

•  Environmental chamber (Control relative humidity to 12 ±3% 

RH and 50 ±3% RH and temperature to 23 ±2° C)

•  Specimen support surface (greater than 10E12 ohms such as 

PMMA, PTFE or polycarbonate)

•  Specimen Pre-Conditioning (23 ±2 degrees C; 3 specimens 

at 12 ±3% relative humidity for up to 48 hours minimum, and 

3 specimens at 50 ±3% relative humidity for up to 48 hours 

minimum

•  Measurement of point-to-point resistance shall be performed 

according to 2.1.2.3 of IEC 61340-4-1 in a minimum of three 

locations for point-to-point resistance, and a minimum of one 

location per square metre for large areas for qualification 

testing, for surface-to-EPA ground resistance. Care shall 

be taken to include, when applicable, resistance between 

separate tiles.

•  Reporting Test Results

  •  Minimum, median and maximum readings for both 

resistance-to-ground and point-to-point resistance in ohms 

at low relative humidity

  •  Minimum, median and maximum readings for both 

resistance-to-EPA bonding point and point-to-point 

resistance in ohms at moderate relative humidity

  •  Temperature

  •  Relative humidity

  •  Actual duration of conditioning

  •  Test equipment used

Periodic Audit of Installed Product Test 
Procedure Guideline

Measure Rg Resistance to Ground

 

Test Procedure:

•  The material shall be tested without cleaning of the surface 

with liquid cleaning agents. Loose dust may be removed by 

gentle brushing or by blowing with clean, dry air.

•  Remove from the surface only those items that might 

interfere with the test. 

•  ESD sensitive devices shall also be removed 

•  Clip the sensing test lead with shielded plug to grounded 

EPA bonding point

•  Use one 2.27 kg Electrode on other test lead and place 

Electrode the farthest convenient point on the surface

•  Press button and hold Test Button until measurement is 

displayed

•  Perform additional measurements placing Electrode on the 

most commonly used or most worn area

•  Per EN 61340-5-1  paragraph 10.1 “All audit measurements 

shall be carried out under the normal operating environment, 

and the temperature and relative humidity shall be recorded”

If measurement is outside acceptable limits, clean surface & 

retest to determine if cause of failure is insulative dirt layer 

or the ESD protective product.  Note: Use an ESD cleaner 

containing no insulative silicone (i.e. Reztore™ Antistatic 

Surface and Mat Cleaner)

Summary of Contents for 99105

Page 1: ...ance Measurement Test method A 1 Floor working surface or storage rack working surface A 2 Seating A 3 Garments A 5 4 Gloves finger cots and tools Neither 20 mm plates nor clip connectors provided Versatility Measures Rp Rg or Resistivity requires optional Resistivity Attachment item 99107 Quick Checks Surface resistance exponent number illuminates immediately i e 8 10E8 ohms or 100 000 000 ohms M...

Page 2: ...ponent or power of the number is displayed immediately with LED illuminating number If 8 is illuminated the measurement is in the 10E8 or 10E8 ohm range To make a measurement the button is to be depressed for 15 20 seconds for the unit to make numerous readings and calculations Testing in accordance with EN 61340 5 1 requires 15 to 20 seconds of electrification in contrast most analog type meters ...

Page 3: ...ectrode on different sections Clean surface for Laboratory test but do not initially clean surface for installed products if fails clean and retest Laboratory Test Procedure Guideline For laboratory test of ESD Working surfaces Floor Materials Footwear Garments or Seating best advice is to follow test methods of Annex A of EN 63140 5 1 The test specimens and electrodes should be cleaned twice with...

Page 4: ...cally perform a lot sampling at incoming inspection Floor Per EN 61340 5 1 paragraph 10 1 These audits shall be performed at intervals not exceeding those defined by the ESD co ordinator Users typically perform these quarterly Seating Per EN 61340 5 1 paragraph 10 1 These audits shall be performed at intervals not exceeding those defined by the ESD co ordinator Users typically perform these quarte...

Page 5: ...ate to 2 5 except at 10E11 and 10E12 5 Thermometer accurate to 1 F Humidity meter accurate to 2 Test leads 99 Isopropyl alcohol and cleaning wipes SET UP A Test Area Area needs to be free of any high voltage transformer or power supply Not under any type of fluorescent lighting or high power lighting B Worksurface needs to be covered with conductive mat at 1 0 x 10E3 or less connected to earth gro...

Page 6: ...d or replaced at Charleswater s option free of charge Call Customer Service at 00 44 0 1892 665313 for a Return Material Authorisation RMA and for proper shipping instructions and address You should include a copy of your original packing slip invoice or other proof of purchase date Any unit under warranty should be shipped prepaid to the Charleswater factory Warranty replacements will take approx...

Reviews: