Part 7:
SOLID STATE CIRCUIT BOARDS
7.4.1 Timer Control and Time Select (Fill/Chemical Dispenser) Board
Connector Descriptions
Refer to Fig. 7.4.1 on page 72 and the test point descriptions below.
Timer Control Board Connectors:
CN1 —
10 pin connector accepts the cable from the Time Select Board.
There are no test points on CN1.
CN2 —
8 pin connector accepts the touchpad display cable.
Refer to Section 7.4.3, page 76, for test points.
CN3 —
10 pin connector hard-wired to the circuit board.
There are no test points on CN3.
CN4 —
4 pin connector accepts the status LED cable from the touchpad/display.
There are no test points on CN4.
CN5 —
4 pin connector accepts the wiring harness connector.
120VAC test points for:
• Wash pump output
• Fill/Rinse valve
• Drain pump and drain valve
CN6 —
4 pin connector accepts the wiring harness connector.
120VAC test points for:
• AC Power input to the Timer Control Board
• Sanitizer pump output
• Rinse aid pump output
• Detergent pump output
CN7 —
4 pin connector accepts the wiring harness connector. (UH-200B, UH-200 only)
Remove harness connector and connect jumper between Common
and LOW, MIDDLE, or LIMIT pin on circuit board to to check water level LED’s.
Associated LED should illuminate on board when pins are jumped.
T1,T2—
Spade terminals on board.
120VAC test points for power input
• T1 accepts AC Hot harness wire from Temperature/Pressure Display Board
to power water level control transformer (UH-200B, UH-200 Only) and
the wash pump, fill/rinse valve, and drain pump/drain valve (All Models).
• T2 accepts AC neutral harness wire (All models)
T3,T4—
Spade terminals on board accept wires from door safety switch.
Time Select (Fill/Chemical Dispenser) Board Connectors:
CN1 —
10 pin cable to the Timer Control Board. Connector hard-wired to board.
There are no test points on CN1.